5. Operation
System settings (continued)
The RUN OPTIONS function displays the options with which a
program is started.
The green check mark indicates that the function has been
enabled.
Pre-test:
• Ifthisoptionisenabled,aretortllinganddrainingtest
is carried out with the rst reagent in the program before
the program is started. This is to ensure that the system is
functioning properly.
The congured option of a respective program always applies to
ALL inltration programs!
Printer:
if this function is selected, a program report is generated after the inltration is complete and can be saved as a PDF le.
The report contains the following data on the program that has run:
• Theprogramname
• Theprogramnumber
• Theinltrationprogramthatwasexecuted
• Runlogeventsthatoccurredduringtheprogramexecution.
The program reports can be saved to a USB stick using the SAVE PDF PRINTOUTS in the ENTRIES menu.
Fig. 94
Select RUN OPTIONS
If a program has been selected or is already running, the run options can also be changed immediately at
any time in operator mode by pressing the RUN OPTIONS button in the graphic program window.
Make the change and conrm it with OK in the window that opens.