Manual
71
Leica ASP6025
5. Operation
Fig. 89
The temperature of the three paran baths and the paran station can be dened in the
INSTRUMENT eld.
• PressPARAFFIN BATH SET TEMP. to open the entry window (Fig. 90) for the paran
bath temperature.
• Youcanselectaparanbathtemperaturebetween50°Cand65°C.Theinputvalueis
paran-specic and depends on the melting temperature of the selected paran type.
• EnterthedesiredparanbathtemperatureandthenpressOK to conrm.
System settings (continued)
INSTRUMENT
Fig. 90
Here you can select the maximum number of baskets with which the retort can be loaded.
Depending on this selection, the retort is lled with the respective number of reagents during
the program ow.
The load options that can be selected are:
• 2 baskets (corresponds to 3.8 l lling) or
• 3 baskets (corresponds to 5.0 l lling).
The corresponding symbol is displayed in the status bar.
The green check mark indicates the active level control.
Fig. 88
Select BASKET MODE
You should always set the value to the highest permitted temperature for
theparanusedsincethereisasmalltemperaturedropwhentheretort
is being lled.