Manual
63
Leica ASP6025
4.8 Switchingotheinstrumentcorrectly
If the instrument must be completely switched o or disconnected from the power supply, please
proceed as follows:
• PresstheBACK button to switch to a window that contains the EXIT button.
• Apromptappearsaskingifyoureallywanttoshutdownthesystem(exit)(Fig. 70).
Pressing YES saves all current data and shuts
down the instrument.
The instrument must then be switched o using the ON/OFF switch on the rear panel of the instrument
(Item 37 in Fig. 3 / Fig. 26).
4. Instrument Setup
Fig. 70
Pressing the NO button leads back to the
original program window.
37
Caution!
TheASP6025isallowedtobeswitchedocompletelyinthiswayonly.Otherwise,severedamageto
the instrument's hardware and loss of data can occur.
Afteracompleteshutdown,theparancoolsdownandbecomeshard.Youmaynolongerbeableto
removetheparanbathdrawersfromtheinstrument.
Important!
IftheASP6025istoremainshutoforalongperiodorbetransported,itismandatorytocarryouta
SMARTcleaningbeforeitisswitchedo.(Formoreinformation,refertoChap. 6.1.4)