Manual
56
Instructions for Use V 2.2 RevI – 08/2013
4. Initial Operation
4.6 Alarm functions
If a user action on the instrument is required, an informational
message is rst displayed on the screen (Fig. 61) and an acoustic
alarm is emitted. This message describes what must be done and
provides further information about the situation.
When the user conrms this message by pressing the YES but-
ton, this message disappears from the screen. No other alarm
is triggered.
If the user does not confirm the message within the specified time by pressing the YES key, an instrument
alarm is triggered. This instrument alarm is an acoustic signal (sound le) that can also be played back cyclically.
This setting (the cyclic repetition of instrument alarms) is congured by the users and is valid for all messages.
The user can also congure the repetition time.
When the instrument alarm is conrmed with the YES button, the alarm sound is switched o and the message
disappears from the screen. No other alarm takes place (no local alarm and no remote alarm).
The alarm sound can also be disabled using the SUSPEND ALARM button. This disables the alarm sound, but
the message remains displayed on the screen. If the user then does not conrm the message within a certain
time by pressing the YES key, the alarm sound is output once again.
Fig. 61
Informational messages
Instrument alarm
The Leica ASP6025 is equipped with 3 dierent acoustic alarm functions:
In the ASP6025, states can occur that require the attention of, or a decision on the part of, the user. In the sim-
plest case, these are conrmations for continuing in the inltration program. In addition, however, during con-
tinuous monitoring of the hardware, errors can be identied that must be eliminated as soon as possible for a
running inltration program to be nished successfully. Correspondingly, all messages are classied according
to their severity.