Manual

53
Leica ASP6025
Labels for RTU-bottles (Fig. 56)
10
Fig. 56
The labels for the RTU-bottles in the drawer come in the same
eight colors as for the system bottles.
The label has lling level marks for the RTU-bottles printed on it,
therefore it must be applied with particular care.
Fig. 57
73
Detail 1
Detail 2
4. Instrument Setup
To ensure that the lling level marks are valid, apply the narrow
side of the label exactly at the point on the RTU-bottle (10)
where the bottleneck ends (Detail 1 in Fig. 57) then press it on
vertically from top to bottom.
The symbol at 5 l (Detail 2 in Fig. 57) means that the RTU-
bottle must be lled up to that point if the retort is to be lled
up to the third level sensor (corresponds to loading with three
baskets).
Accordingly, the symbols at 3.8 l indicate the fill volume for
lling the retorts up to the second level sensor.
The station number can likewise be noted on the label as well
as the date when the specied reagent was commissioned
(Fig. 57).
This must also be adjusted in the reagent status for a
newly lled RTU-bottle.
Caution!
The specied lling level marks (5.0 l and
3.8 l) are valid only for the RTU-bottles includ-
ed in the delivery package. If other contain-
ers are used, the specied ll volumes on the
label will not agree.