4.4.8 Reagent cabinet with system bottles
• Theninewhiteplasticsystembottles(15) are located to-
gether with the blue condensate bottle (14) in the reagent
cabinet under the three paran baths (Fig. 49).
Fig. 49
4. Initial Operation
1514
Fig. 50
When a reagent has been assigned a system bottle, this is
displayed on the screen.
Fig. 51
• Allsystembottles(15) have a max. capacity of 5 l .
The ll levels for lling with 3.8 l and 5.0 l are embossed on
the front of each container.
• Wheninsertingasystembottle,alwaysensurethatthe
retaining ring (35) is tightened and the connection port
(65) clicks into place correctly with the O-ring (65.1).
65.1
65
Regularly grease the O-ring (65.1)
with Molykote grease for sealing rings
(included in the delivery package).
11
12
15
35
• Toaccessthesystembottles,openthetwolowerPlexiglas
doors (11, 12).
• Thesystembottlesaredesignated"S1" to "S9" and are
combined in a separate area on the screen (Fig. 50). The
condensate bottle, designated "C", is on the left.
Detail: Connection of the system
bottle
Only Leica system bottles can be used.
Once the pressure point is exceeded, you can
clearly feel it click into place, thus ensuring an
optimum t in the reagent cabinet.