Manual
49
Leica ASP6025
Pull-out drawer with RTU-bottles (continued)
Height (max.): 350 mm
Height
(min.): 245 mm
Width
(max.): 200 mm
Width
(min.): 155 mm
Depth
(max.): 180 mm
Depth
(min.): 135 mm
4. Instrument Setup
Permitted dimensions for RTU-bottles in the drawer:
Bottle neck diameter
Outer (
max.): 54 mm
Outer (
min.): 38 mm
Inner (
max.): 44 mm
Inner (
min.): 27 mm
• TheRTU-bottlesaredisplayedintheSMART SCREEN in a
separate area on the screen. They are designated "D1" to
"D6" (Fig. 48).
• Additionally,theRTU-bottlescanbeemptiedusingthe
remote drain system.
• Thehosefortheremotellanddrainsystemcanalsobe
stored in the drawer.
Fig. 48
If RTU-bottles other than those installed by Leica are used in the drawer, these must be resistant to the
following reagents and temperatures and dimensionally stable.
For cleaning solutions up to a temperature of 71 °C
For progress reagents up to a temperature of 64 °C
The dimensions must not fall below or exceed those listed below.