4. Initial Operation
Pull-out drawer with RTU-bottles (continued)
Fig. 47
• Thebottomrightofthedrawerhasaspigot(64, Fig. 47),
for draining spilled or overowing reagents neatly. Then,
the drawer should be cleaned.
Spigot closed
Spigot open
64
64
Fig. 46
10
• SevenRTU-bottles(10) are supplied for the drawer with the instrument (Fig. 46)
- 6x for tissue inltration,
- 1x for the cleaning process (extended cleaning program by detergent).
Each container has a screw plug.
• Eachcontainerhasamax.capacityof5l.
Fig. 47a detail
Only RTU-bottles from Leica are approved for this instrument. However, if dif-
ferent bottles are used, the customer must INDEPENDENTLY check that they are
suitable for such use.
(For information about temperature and dimension requirements, see page 49 -
Warning message.)
Important!
Always keep the spigot (Fig. 64 detail) closed.
The spigot (Fig. 47a detail, item 64) may be
opened for cleaning purposes only, as other-
wise ammable reagents could end up on the
heateroftheparanbathsandignite.