Manual
45
Leica ASP6025
4. Instrument Setup
4.4.6 Theparanbaths
• Theinstrumenthasthreeheatedparanbaths(58), each
with a max. capacity of 4.9 l of liquid paran.
• Theyarelocatedbehindaap(57) over the reagent compart-
ment with the system bottles (15).
To access the ap, open the two lower Plexiglas doors (11, 12).
• Theindividualparanbathscanbepulledoutforlling
(Fig. 41) and can also be removed from the compartment
for cleaning.
Fig. 41
11
57
58
15
58
58
• Thebathscanbelledwithliquidparanmanuallyorfrom
the paran station. It is also possible to ll with pellets - the
melting time then is approx. 720 min.
58
Caution!
Never attempt to forcefully remove the wax
baths when the instrument is cold, because
doing so can damage the instrument.
Caution!
Proceed slowly and carefully when pulling out
or removing a paran bath—never jerk it
out.Theparaninsideitisuidandhot—it
can cause burns. The handles and covers are
also hot, therefore always wear gloves and
proceed carefully.