Manual
44
Instructions for Use V 2.2 RevI – 08/2013
Fig. 39
• Theparanstationmustbepulledouttobelled.Ithasacover(55) for
better thermal insulation and as splash protection. The cover has two holes
in it to make it easier to remove.
53
4.4.5 Theparanstation
• Theparanstation(54) is located behind the left door (53) of the instru-
ment, below the retort (Fig. 37). It ensures that sucient fresh, liquid
paran is always available for the paran baths. The capacity is 5.0 l of
liquid paran.
• Therearetwomarksintheinteriorofthestationthatdisplaytheminimum
ll level when lled with pellets or liquid paran (Fig.38). The level must
not fall below these marks.
56
55
4. Initial Operation
Fig. 37
56
Detail:
Sieve in the
paranstation
54
Fig. 38
Top mark:
Minimum ll level when lling
with pellets for melting.
Bottom mark:
Minimum ll level when lling
withliquidparan.
Fig. 40
• Theparanstationcanbelledwithparanpelletsorliquidparan.
If lled with pellets, the melting time is approx. 6.0 h.
• Checkandcleanthesieve(56) monthly.
Iftheparanstationhasbeenpulledout,thisisindicated
by a red border (Fig. 40) on the SMART screen.
The indicator light on the lower right of the icon will also
be red.
Proceed slowly and carefully when pulling out the
station—neverjerkitout.Theparaninsideitisuidand
hot—it can cause burns. The cover is also hot, therefore
always wear gloves!