Instructions for Use V 2.2 RevI – 08/2013
Fig. 39
• Theparanstationmustbepulledouttobelled.Ithasacover(55) for
better thermal insulation and as splash protection. The cover has two holes
in it to make it easier to remove.
53
4.4.5 Theparanstation
• Theparanstation(54) is located behind the left door (53) of the instru-
ment, below the retort (Fig. 37). It ensures that sucient fresh, liquid
paran is always available for the paran baths. The capacity is 5.0 l of
liquid paran.
• Therearetwomarksintheinteriorofthestationthatdisplaytheminimum
ll level when lled with pellets or liquid paran (Fig.38). The level must
not fall below these marks.
56
55
4. Initial Operation
Fig. 37
56
Detail:
Sieve in the
paranstation
54
Fig. 38
Top mark:
Minimum ll level when lling
with pellets for melting.
Bottom mark:
Minimum ll level when lling
withliquidparan.
Fig. 40
• Theparanstationcanbelledwithparanpelletsorliquidparan.
If lled with pellets, the melting time is approx. 6.0 h.
• Checkandcleanthesieve(56) monthly.
Iftheparanstationhasbeenpulledout,thisisindicated
by a red border (Fig. 40) on the SMART screen.
The indicator light on the lower right of the icon will also
be red.
Proceed slowly and carefully when pulling out the
station—neverjerkitout.Theparaninsideitisuidand
hot—it can cause burns. The cover is also hot, therefore
always wear gloves!