Manual

39
Leica ASP6025
4.4.1 The retort
4. Instrument Setup
Retort locking
• Theretortlocksautomaticallywhenaprogramisstarted.
• Unlockitbypressingontheunlockingmechanismswitch
(8.
2) below the screen (Fig. 29).
• Amessageboxappearsonthescreen;YES must be entered
here to conrm unlocking of the retort while a process is in
progress.
• WhentheYES button is pressed (Fig. 29), the reagent vapors
are removed by suction, ltered and the uid level in the
retort is lowered. Then the pin (43) unlocks the handle (8.1)
and the retort can be opened.
• PressingNO continues the process.
Fig. 28
43
• Tocontinuetheprocessafteropeningtheretort,movethe
handle back into the locking position (Fig. 28). Then press
the CONTINUE button on the screen.
8.1
• Toopentheretort,twistforwardtheblackhandle(8.1) on
the retort lid (4) (arrow in Fig. 27). The lid opens upwards.
Fig. 27
8.1
4
Fig. 29
8.2
Locked
Open
Pin (43) can also be unlocked manually - e.g.
after a power failure. To do so, push down the
pin using the Allen key provided.
Keep your distance when opening the retort
lid, particularly if the reagents are heated.