3. Instrument Components and Specications
3.4.3 Instrument specications
• TheLeicaASP6025isamodulartissueprocessorwithaReagentManagementSystemopti-
mizedbyanintegratedmeasurementsensor(ethanolconcentration)thatsupportsconsistently
high specimen quality and helps to reduce reagent consumption.
The concentrations of all ethanols and isopropanols located in the processor are measured and
displayed in the REAGENT STATUS submenu.
• TheLeicaASP6025canbeoperatedusingcustomer-specicorpreinstalled,validatedinltra-
tion programs.
• Forthispurpose,thereare13preinstalled,noneditableprocessingprogramsavailabletothe
user. These include 3 autorotation, 5 xylene and 5 xylene-free programs.
• 20processingprogramsarefreelyconfigurablewithupto15steps(temperature;time;
reagent;threepressure/vacuumoptions).
• Thequickstartsystemenablesyoutostarteachinltrationprogramimmediatelyfromthe
FAVORITES window (max. 10).
• The"time-optimized"inltrationprogramshelptoincreaseproductivityinthelabbysigni-
cantly shortening inltration times. They either include xylene as the intermedium or they are
xylene-free.Inthelattercase,hazardousxyleneisreplacedwithisopropanol.
• Withthepreinstalledautorotationprograms,theethanolconcentrationismeasuredauto-
matically and when the concentration falls below a threshold value a message is displaying
indicating that it is time to change the used ethanol. When the used ethanol is replaced,
the fresh ethanol added to the system is always undiluted (100 %) ethanol. This means that
time-consuming dilution and the errors that can occur during that process, as well as contact
(inhalation!) with the solvents are entirely avoided.