3. Instrument Components and Specications
3.2 Technical data (continued)
Hardware and Software:
• ColorLCDtouchscreen.
• User-friendly,intelligentsoftware.
• ThreeUSBports.
• Alarmsystemwithtworemotealarmsockets.
• Password-protectedinstrumentsupervisormode.
• Multiplespecimenprotectionsystem.
Capacities:
• 20freelycongurableprograms,eachconsistingofupto12reagentand3paranprocessing
steps.
• Timeperprogramstep:0to23hours,59minutes.
• Delaytime:max.6days
• Upto300cassettescanbeprocessedsimultaneously.
• Threecleaningprogramsfortheretort.
• 9internalsystembottles.
• 6RTU-bottlesinthedrawer.
• 3paranbaths.
• 1paranstationformeltingparan.
• 1condensatebottle.
• Reagenttemperatureselectablefrom35°Cto60°C
or room temperature, depending on the reagent.
• Parantemperatureselectablefrom50°Cto65°C.
• Cleaningtemperatureof62°C(ethanol)to67°C(R-xylene)
• Upto100reagentnamesinmemory.