3. Instrument Components and Specications
3.2 Technical Data
Nominal supply voltages: Two factory-preset voltages (not user-adjustable):
120 V + 10 % or
230 V + 10 %
Nominalfrequency: 50to60Hz
Main fuses: 2 melting fuses, 32 x 6.3 mm, UL-approved
• for120V;T15A125VAC
• for230V;T10A250VAC
Nominal power: 1700 VA for both versions
Dimensions, (L x W x H), in mm: 680 x 750 x 1500
Empty weight, unpacked: max. 210 kg
Weight, packed: max. 235 kg
Operating temperature range: +15 °C to +40 °C
Storage temperature range: +5 °C to +50 °C
Relative humidity: 10 % to max. 80 % non-condensing
IEC 1010 classication: Protection class 1
Pollution degree 2
Overvoltage installation category II:
Ambient pressure: 740 to 1100 hPa
Operating elevation: up to a max. of 2000 m above sea level
A-weighted noise level: < 70dB (A)
Local/remote alarm relay: 30 V DC, max. 1 A
120 V AC, max. 0.25 A
2 terminals for 6.3mm stereo jack.
Each with isolated switching contact
(operable both as normally-open and
normally-closed circuit)
Parancontainers
Paran melting station: 1
Number of paran baths: 3
Capacity: 4.9l per paran bath
Melting time: in paran station: max. 6.0 h, in paran bath max. 12.0 h
Temperature: 50 to 65 °C (+ 6 K - 0 K)