170
Instructions for Use V 2.2 RevI – 08/2013
7. Troubleshooting
7.3 Typical ll or drain problems
Fill and drain problems may be due to several causes:
1stThereisinsucientreagent
• Verify that the ll level in the system bottles, RTU-bottles and paran containers is
sucient.
2. Insucientpressureorvacuum
The instrument is unable to create sucient vacuum (for a ll) or pressure (for a drain) in the
retort.
• Check whether the system bottles and RTU-bottles concerned have been inserted
correctly.
• Verifythatallsystembottlesarefullypushedhomeintotheirsockets.
• Checkwhetherdirtdepositshavecollectedundertheretortseal.
3. There is a blockage in the air or reagent lines
Blockages in the reagent lines are typically caused by paraffin or tissue debris. If it is not
possible to drain the reagent in the retort, try to remove the blockage as follows:
• Heattheretorttothemaximumpossibletemperature(consistentwiththereagentcur-
rently in the retort, if any). Leave the retort at this temperature for at least 15 minutes.
• Whentheretortreachesmaximumtemperature,attempttocarryoutallinganddrain-
ing cycle.
• Ifthereisnoreagentintheretort:
Clean the retort in a cleaning program (see Chapter 6.1).
If you cannot determine or eliminate the problem by carrying out the above
listed steps, call Leica Technical Service before attempting to run any further
programs.