Manual
167
Leica ASP6025
7. Troubleshooting
• Ifno inltration program is running when a power failure is detected, no further actions / backup measures are taken -
the instrument shuts down automatically.
When a program is running:
• Theprogramthatisrunningisstopped,regardlessofthecurrentstatusofprocessing.
Fig. 239
• Ifareagentisbeingusedforwhichnosafereagentisinthe
instrument when the program starts, a warning message
appears (Fig. 239). However, the program launch is NOT
aborted.
• Ifasafe reagent is in the retort, the reagent can be kept in
the event of power failure, though the retort may have to be
lled to a high enough level so that no tissue sample is dry.
This operation is automatically controlled by the instrument
and cannot be inuenced by the operator.
• Ifno safe reagent is in the retort, depending on the cur-
rent status of processing the program that is in progress, a
"decision" is made about how to proceed.
Current step for power failure Safeguard in safe reagent
Formalin (xation) Formalin
Ethanol (dehydration, diluted) Formalin
Ethanol (absolute) Formalin
Xylene (intermedium) Xylene (heating o)
Isopropanol (intermedium) Formalin
Paran Paran (heating o)
ParaLast ParaLast
Toconservethebattery,allheaters,theagitatorandthemonitorareswitchedodirectly.