6. Cleaning and Maintenance
Replacing the active carbon lter
• Thelifeoftheactivecarbonlterwilldependonthereagent
types used and how frequently they are used in the instru-
ment.
• Thelter(13) should be replaced at least every 30 days
(Fig. 227).
After replacing the lter, open the SYSTEM MONITOR menu
to set the lter replacement date to the current date. See also
Chapter 5.1.4.
Fig. 227
Emptying the condensate bottle
• Thebluecondensatebottlemustbetakenout,emptiedand
reinserted at regular intervals (weekly). This should be done
sooner if a prompt to that eect is issued by the software
(Fig. 225).
Fig. 225
Filter label
14
13
Fig. 226
Dispose of waste solvents with care according
to local regulations and the waste manage-
ment policy of the company or institution.
When inserting the new lter, make sure that
it is pushed all the way into the compartment
and that the ap is closed.