Manual

14
Instructions for Use V 2.2 RevI – 08/2013
2. Safety
Warnings – Cleaning and maintenance
Material safety data sheets can be obtained from the supplier of the chemicals.
Alternatively, they can be downloaded from the following website:
http://www.msdsonline.com
• Beforeanymaintenance,switchotheinstrumentandunplugitfrompowersupply.
• Donotusesolventsthatcontainacetoneforcleaning.Noliquidmaybespilledintotheinternalcomponents
of the instrument – neither during operation nor during cleaning.
• Whenusingcleaners,complywiththesafetyinstructionsofthemanufacturerandthesafetyregulationsof
your laboratory!
• Checkthecondensatecontaineratleastonceaweekand,ifnecessary,drainit.
• SystembottlesandLeicaRTU-bottlesmustnotbecleanedinadishwasher-thecontainersareNOTdishwasher-
proof.