Manual
153
Leica ASP6025
6. Cleaning and Maintenance
6.1.3 Cleaning protocols
I. Rapid cleaning program
Step: Reagent: Time (min): Temp. (°C): Pressure/Vacuum:
1 Cleaning xylene 6 67 Ambient
2 Cleaning alcohol 3 62 Ambient
Dry step:
Empty retort
3 Vacuum step 10 67 Vacuum
4 Ventilation step 2 Ambient Ambient
Optional step instead of "Dry step":
5 Cleaning water 1 65 Ambient
II. Standard cleaning program
Step: Reagent: Time (min): Temp. (°C): Pressure/Vacuum:
1 Cleaning xylene 12 67 Ambient
2 Cleaning alcohol 6 62 Ambient
Dry step:
Empty retort
3 Vacuum step 10 67 Vacuum
4 Ventilation step 2 Ambient Ambient
Optional step instead of "Dry step":
5 Cleaning water 1 65 Ambient
III. Extended cleaning program
Step: Reagent: Time (min): Temp. (°C): Pressure/Vacuum:
1 Cleaning xylene 25 67 Ambient
2 Cleaning alcohol 10 62 Ambient
3 Cleaning agent 16 65 Ambient
4 Cleaning water 16 65 Ambient