Manual

151
Leica ASP6025
6. Cleaning and Maintenance
Running cleaning programs (continued)
Fig. 213
Cleaning after completion of an inltration program
If desired, you can select an additional water step for each retort
cleaning program (EXCEPT the full cleaning with detergents).
(See also Chapter 6.1.3).
Fig. 214
• Toselectthewaterstep,presstheWATER YES/NO button so
that a green check mark appears (Fig.212/213). The water
step is always carried out as the third step in the cleaning
program, that is to say it is not run in addition to the cleaning
program shown in Fig.208.
The three cleaning programs are oered au-
tomatically when an inltration program has
been completed (Fig. 211).
In this case, a retort cleaning cycle MUST be
carried out; this function cannot be canceled.
The water step can only be selected if cleaning
water is present in one of the bottles.
In "Auto by RMS" mode, "Cleaning water"
must have been dened for a station.
In "Concentration" mode, the process water is
automatically used as cleaning water.
If the magnetic stirrer was removed from the
retort, it must be reinserted before the start
of cleaning.
Otherwise, the cleaning reagents in the retort
willnotbeheatedeectively.
If cassettes/baskets are not removed immediately after a paraf-
n run, paran can drop into the previously emptied retorts.
After an inltration program, it is not possible
to avoid cleaning - no CANCEL key is present.
Cleaning MUST be carried out!
Fig. 212