6. Cleaning and Maintenance
Running cleaning programs (continued)
Fig. 213
Cleaning after completion of an inltration program
If desired, you can select an additional water step for each retort
cleaning program (EXCEPT the full cleaning with detergents).
(See also Chapter 6.1.3).
Fig. 214
• Toselectthewaterstep,presstheWATER YES/NO button so
that a green check mark appears (Fig.212/213). The water
step is always carried out as the third step in the cleaning
program, that is to say it is not run in addition to the cleaning
program shown in Fig.208.
The three cleaning programs are oered au-
tomatically when an inltration program has
been completed (Fig. 211).
In this case, a retort cleaning cycle MUST be
carried out; this function cannot be canceled.
The water step can only be selected if cleaning
water is present in one of the bottles.
In "Auto by RMS" mode, "Cleaning water"
must have been dened for a station.
In "Concentration" mode, the process water is
automatically used as cleaning water.
If the magnetic stirrer was removed from the
retort, it must be reinserted before the start
of cleaning.
Otherwise, the cleaning reagents in the retort
willnotbeheatedeectively.
If cassettes/baskets are not removed immediately after a paraf-
n run, paran can drop into the previously emptied retorts.
After an inltration program, it is not possible
to avoid cleaning - no CANCEL key is present.
Cleaning MUST be carried out!