Leica ASP6025
5. Operation
5.6.3 Xylene-free inltration programs
Reagent positioning/Replacement cycle
for time-optimized xylene-free programs:
System bottle S1 contains 70 % ethanol.
System bottle S2 contains 85 % ethanol.
Each program also requires an additional step with ParaLast to reduce reagent carry-over.
The pressure/vacuum cycle requires a minimum step duration of 6 minutes!
Bottle Reagent
S1 70 % ethanol
S2 85 % ethanol
S3 -empty-
S4 80/20 ethanol/isopropanol
S5 80/20 ethanol/isopropanol
S6 100 % isopropanol
S7 100 % isopropanol
S8 100 % isopropanol
S9 - empty -
D1 Formalin
D2 -empty-
D3 ParaLast
D4 -empty-
D5 Cleaning xylene
D6 Cleaning alcohol
Xylene-free inltration programs cannot be renamed, edited or deleted by the user.
To start a xylene-free program, it must have been saved in the list of "FAVORITES". See also Chapter 5.5.3.
The specied isopropanol concentrations are the starting concentrations that apply for the installation of the
ASP6025 modular tissue processor.