5. Operation
5.6 Permanently installed inltration programs
• Thereplacementcycleisdenedbytheethanolconcentrationthresholdvaluefortherstethanolmeasuredbythe
density sensor.
The ethanol threshold value has been set to 55 % by the manufacturer and can be adjusted in 1 % increments between
50 and 60 %.
• Theethanolconcentrationsshownaretheinitialconcentrationsthatapplyforinstallationofthetissueprocessor.
Reagent positioning by autorotation programs
5.6.1 Programs with automatic ethanol rotation
All of the permanently installed inltration protocols described in this chapter have to be validated before use,
i.e. tissue processing using patient tissue for diagnostics, by the laboratory itself based on tissues that are not
intended for diagnostics and in accordance with local or regional accreditation requirements.
Autorotation programs cannot be renamed, edited or deleted by the user.
To start an autorotation program, it must be saved in the list of "FAVORITES". See also Chapter 5.5.3
Bottle Reagent
S1 70 % ethanol
S2 80 % ethanol
S3 95 % ethanol
S4 100 % ethanol
S5 100 % ethanol
S6 100 % ethanol
S7 Xylene
S8 Xylene
S9 Xylene
D1 Formalin
D2 Water
D3 Replacing xylene
D4 Replacingethanol->100%
D5 Cleaning xylene
D6 Cleaning alcohol