Manual

121
Leica ASP6025
5. Operation
When connecting the hose, be sure to press it
onto the O-rings of the drain opening as far as
it will go.
The paran drain hose must be securely
hooked into the external receptacle and
remain there through the entire drainage
process.
When drainage is nished, the hose is
cleaned automatically using air.
Do not remove the hose from the external
receptacle until this cleaning step is com-
plete.
57
• UsingtheSMART SCREEN button, call up the SMART
FUNCTIONS screen (Fig. 175).
Fig. 174
5.4.3 Drainingaparanbath
• Connecttheparandrainhose(36) to the spout
(57) of the paran drain on the front of the instrument
and guide it into a catch tank (Fig.174).
36
Fig. 175
AlwaysdrainaparanbathusingtheRemote
Drain function only. When doing so, you
should use a suitable vessel with a capacity of
atleast10literstodraina5-literparanbath
to the outside (danger of splashing!).