Manual
121
Leica ASP6025
5. Operation
When connecting the hose, be sure to press it
onto the O-rings of the drain opening as far as
it will go.
The paran drain hose must be securely
hooked into the external receptacle and
remain there through the entire drainage
process.
When drainage is nished, the hose is
cleaned automatically using air.
Do not remove the hose from the external
receptacle until this cleaning step is com-
plete.
57
• UsingtheSMART SCREEN button, call up the SMART
FUNCTIONS screen (Fig. 175).
Fig. 174
5.4.3 Drainingaparanbath
• Connecttheparandrainhose(36) to the spout
(57) of the paran drain on the front of the instrument
and guide it into a catch tank (Fig.174).
36
Fig. 175
AlwaysdrainaparanbathusingtheRemote
Drain function only. When doing so, you
should use a suitable vessel with a capacity of
atleast10literstodraina5-literparanbath
to the outside (danger of splashing!).