Leica ASP6025
• Theentireoperationisdisplayedgraphicallyonthescreen
(SMART FUNCTIONS view) (Fig. 165).
• Thereagentispumpedoutoftheexternalcontainerand
initially into the retort. Then, the selected system bottle is
lled.
Fig. 165
Fig. 166
Fig. 167
• Whentheselectedstationislledwithreagent,amessage
window appears (Fig. 166), with a message to the eect that
the hose for remote lling/draining is being blown out with air.
• Thereforeleavethehoseintheremotestationuntilthe
process has ended, or insert it into another suitable container.
• Aftertherinsinghasbeencompleted,amessageappears
(Fig. 167), informing you that remote lling has been com-
pleted. To end the process, press OK.
Draining/lling reagents (continued)
5. Operation