Manual

116
Instructions for Use V 2.2 RevI – 08/2013
Remote lling
Select station Press button
Fig. 162
Draining/lling reagents (continued)
Fig. 163
Fig. 164
• Afteryouhaveselectedareagent,ascreenmessageappears
prompting you to make the necessary preparations.
• Wheneverythingisfinished,pressYES to start filling
(Fig. 164).
5. Operation
• Onthescreen,selectanemptysystembottle(e.g.S6) and
the external receptacle (Remote Station, Fig. 162).
• ThenpresstheSTATION REFILL button.
• Nowthereagentwithwhichthehighlightedbottleisto
be lled (of course, the reagent contained in the external
station) must be selected (Fig. 163).
Remote lling of system bottles takes place in exactly the same
way as remote draining of RTU-bottles, but in the reverse order
of course.
+
Remote lling is only possible for system bot-
tles. RTU-bottles in the drawer must be lled
manually.