Leica ASP6025
Onthescreen(Fig. 156), select the system bottle or
RTU-bottle that is to be drained and then select the recep-
tacle (external station).
Then press STATION DRAIN.
Select station Press button
Draining/lling reagents (continued)
Fig. 158
Fig. 159
Fig. 161
• Amessagewindowonthescreenshowsthattheprocesshas
been completed (Fig. 161). To continue working, conrm this
with OK.
• Ifthesamereagentwillnotbeusedinthenextoperating
step, the retort may have to be cleaned with a cleaning
program.
• Apromptappearsforcarryingoutthenecessarypreparations
(Fig. 159). This must be conrmed with YES.
• ThereagentispumpedoutoftheRTU-bottleintotheretort.
• Oncethistaskhasbeencompleted,theretortisdrainedinto
the external station (79 in Figs. 156 and 157).
• Theentireprocessisdisplayedgraphicallyonthescreen
(SMART SCREEN view) (Fig. 153, right).
Fig. 160
5. Operation
+