Manual
114
Instructions for Use V 2.2 RevI – 08/2013
Remote draining
• Connectthehoseforremotelling/draining(70) and pass
it into the external station (79) (e.g., a bulk container) as
shown in Fig. 157.
Press the SMART SCREEN button.
Fig. 156
Fig. 157
5. Operation
Draining/lling reagents (continued)
The SMART SCREEN (Fig. 156) pops up.
The smart screen is the initial screen for
manual operation of the instrument.
79
79
70
Important!
Dispose of waste solvents with care according
to local regulations and the waste manage-
ment policy of the company or institution.
ATTENTION VERY IMPORTANT!
During remote lling or draining, always
select ONLY the external station (79) and the
bottle that is to be lled or drained in the
SMART FUNCTIONS window.
The retort cannot be lled in the meantime
for this procedure. If necessary, the retort can
be lled from the outside - afterwards the re-
agent has to be emptied TO THE OUTSIDE.