Manual
113
Leica ASP6025
Fig. 155
Hose for remote lling/draining of reagents
• Connectthehoseforremotelling/draining(70) as shown
in Fig. 155.
• Whenconnectingthehosetotheconnectionspout(item
18 in Fig. 155) for reagents, the connection piece (71) must
engage with the spout with a clearly audible click.
18
18
70
71
Remote replacement of reagents
70
5.4 Reagent handling 5.4.1 Filling/draining reagents
5. Operation
The following steps should only be carried out
by trained laboratory personnel, experienced
in handling reagents.
Always wear rubber gloves and safety gog-
gles when handling the reagents used in this
instrument.
Reagents used for tissue inltration can be
both toxic and/or ammable.
When lling or draining a system bottle or
RTU-bottle, ensure that the remote ll/drain
hose is securely hooked into the external sta-
tion and is not removed until the operation is
fully completed. Finally, pressurized air is used
to clear the hose after each ll / drain so that
no dirt remains. The hose must therefore not
be taken out of the external station until this
cleaning step is complete.
• Theremotedrainingorfillingfunctionispossiblein
3.8 and 5 liter mode. - Both system bottles and RTU-bottles
can be lled or drained.