Manual

99
Leica ASP6025
The RMS system – adjusting warning threshold values (continued)
Tables with warning threshold values
5. Operation
Xylene protocols
Reagent recommended replacement
Formalin after 600 cassettes
70 % ethanol after 600 cassettes
85 %/90 % ethanol after 1500 cassettes
1 100 % ethanol after 1500 cassettes
1 xylene after 1500 cassettes
1 paran after 1800 cassettes
Cleaning xylene after 6 cycles
Cleaning alcohol after 6 cycles
Cleaning water after 6 cycles
Xylene-free protocols
Reagent recommended replacement
Formalin after 600 cassettes
70 % ethanol after 600 cassettes
85 % ethanol after 1500 cassettes
1 (ethanol/isopropanol) 80/20 after 1500 cassettes
1 isopropanol 100 % after 1500 cassettes
ParaLast after 2400 cassettes
1 paran after 2400 cassettes
Cleaning xylene after 6 cycles
Cleaning alcohol after 6 cycles
Cleaning water after 6 cycles