7. Cleaning and Maintenance
Fig. 87
7.2.2 Periodic cleaning and maintenance (continued)
44
Fig. 88
Fig. 89
Use the cleaning tool together with the microfiber cloth (Fig. 87) as shown in Fig. 89.
45
46
47
The level sensors must be cleaned of all residues (reagents and paraffin) after every
cleaning program!
Youwillbepromptedtocleanthesensorsbytheinstrumentsoftwareattheendofeach
cleaning program (Fig. 88, 44).
NEVER use the cleaning tool WITHOUT the accompanying microfiber cloth, as other-
wise the sensors will be scratched!
The cleaning tool has two different spoons. The larger spoon (Fig. 87, 46) is for use
with the ASP6025, the smaller (Fig. 87, 47) is for use with the ASP200 S / ASP300 S.
To prevent damage to the sensors, NEVER use the larger spoon for cleaning the sen-
sors in the ASP200 S / ASP300 S.
Cleaning the level sensors