User guide

67
Leica ASP200 S
5.4 Reagent status
This window (Fig. 77) is used to display and update the current reagent
status of the stations. In addition, reagent stations can be marked as
full or empty here.
Fig. 77
The following is displayed:
• Thecurrentageofeachreagent
• Thecurrentstatusofeachstation
• Criteriathattriggerwarningmes-
sages
(fields highlighted in red)
To change the status of a reagent,
highlight the corresponding row and
press the button symbol for the re-
spective action.
In addition, you can switch directly
to the REAGENTS and STATIONS
windows to make current changes.
Reagent warning messages
If the RMS is active, warning messages are out-
put if reagents are used for too long.
The output of the warning messages can take
place after up to 3 criteria for normal reagents
and after up to 6 criteria for paraffins.
Warning messages are displayed at the end of a
cleaning program. Then, you can switch directly
to the REAGENT STATUS display.
Warning messages also appear (with a yellow
background) in the windows in which programs
are started.
The criteria for warning messages
For normal reagents, it is the number of the fol-
lowing since the last reagent change:
• Editedblocks
• Runprograms
• Dayspassed
For paraffins, warning messages can also be
output according to the number of the following
since the last paraffin cleaning:
• Editedblocks
• Runprograms
• Dayspassed
5. Operation