User guide
57
Leica ASP200 S
5. Operation
Remote draining
• Onthescreen,selectthereagentcontainer
(No. 3) and the receptacle (External).
Then press STATION DRAIN.
• Checkthatnoreagenthasleaked.
Remote filling
• On thescreen,selectanemptyreagent
container (No. 3) and the external receptacle
(External).
Then press the STATION REFILL button.
• Theremotefillingshouldbecompletedinless
than 170 seconds.
• Checkthatnoreagenthasleaked.
Select station Press button
Filling/draining reagents
From the FAVORITES screen, press MENU. In
the MENU FUNCTIONS window, tap the SMART
SCREEN button.
The SMART SCREEN (Fig. 55) pops up. The smart
screen is the initial screen for manual operation
of the instrument.
Fig. 55
→
Fig. 56
Status bar,
displayed in supervisor mode.
Status bar
The following values are displayed.
• Pressureandtemperatureoftheretort
• Waxbathtemperature
• Valueofthelevelsensoroftheretort
• Retortlid-closedoropen
The following steps should only be
carried out by trained laboratory
personnel, experienced in handling
reagents.