User guide

46
Instructions for Use V 1.4, RevE – 03/2013
Entering / modifying reagent thresholds
If warning thresholds are required for a certain reagent, enter them as fol-
lows:
• Highlightthereagenttobemodified,eitherbypressingthereagentname
or using the UP/DOWN buttons.
• Presstheheaderoftheparametertobechanged;thecorresponding
entry screen appears.
• Enteranewthresholdvalueor-ifnowarningisdesired-pressCLEAR
to remove the threshold altogether.
• PressOK to confirm.
5. Operation
Changeable parameters
Fig. 42
Fig. 43
Allocation of reagents to the correct reagent group is the basis for compatibility monitoring.
Allocation to the wrong reagent group can lead to reagent cross contamination.
The respective threshold value applies for all reagent stations con-
taining the same reagent.
If a reagent is already used in a program, it can neither be renamed
nor can it be allocated to another reagent group!
The corresponding symbols will be disabled (i.e. they will not be
surrounded by a blue border).
If a reagent is renamed, all stations and programs linked to that re-
agent need to be reedited as well!
Changing reagent names or reagent groups
• Highlightthereagentthenameorgroupofwhichyouwishtochange.
• Pressthecorrespondingbuttonsymbolintheheadline.
• Intheentrywindow(orviathekeyboard),enterthenewreagentgroup
allocation / the new reagent name.
• PressOK to save the new reagent group / reagent name.