5.1.4 Editing the reagent list
Adding new reagents
1. Enter the reagent name and allocate the new reagent to the reagent
group it belongs to.
2. Determine how many stations (bottles) work with the new reagent.
3. Fill the stations (bottles) according to the reagent list.
Entering reagent names
Takes place via the REAGENTS menu option.
Youmustbeloggedonatsupervisoraccessleveltoproceed.
• Fromthestartscreen,pressMENU in the MENU FUNCTIONS window;
there, press the MORE button.
• The MORE MENU FUNCTIONS window appears. There, press the
REAGENTS button.
• TheSET UP REAGENTS AND WARNING THRESHOLDS window opens.
• Toaddareagent:
• PressINSERT to display the keyboard.
• Enterthenewreagentname.
• PressOK to confirm.
• Youwillthenautomaticallybepromptedtoselectthereagentgroup:
→
Fig. 40
Select the reagent group
Allocate the new reagent to the desired group
and press OK to confirm.
5. Operation
Fig. 41