User guide

21
Leica ASP200 S
3.5 Compatible reagents
3. Instrument Components and Specifications
Hardware and Software:
• LargecolorLCDtouchscreen.
• User-friendly,intelligentsoftware.
• 3.5"diskdriveandprinterport.
• Alarmsystemwithtworemotealarmsockets.
• Password-protectedinstrumentsupervisormode.
• Built-inmultiplespecimenprotectionsystem.
Capacities:
• 15programsthatconsistofupto10reagentand3paraffinprocessingstepseach.
• Timeperprogramstep:0to99h,59min.
• Delaytime:max.7days
• Upto200cassettescanbeprocessedsimultaneously.
• 4programmablecleaningprogramsfortheretort.
• 1paraffincleaningprogram
• 10reagentcontainers.
• 3paraffinreservoirs.
• 3containersforcleaningsolutions
• 1condensatecontainer
• Reagenttemperatureselectablefrom35°Cto55°Corroomtemperature.
• Paraffintemperatureselectablefrom40°Cto65°C.
• Choiceofthreeretortdrainratesof80,120or140seconds.
• Upto100reagentnamesinmemory.
3.4 Technical data (continued)
Use of the ASP200 S is permitted with the reagents specified in Chap. 3.5 only. These reagents
must be validated before use, i.e. tissue processing with patient tissue for diagnostics, by the
laboratory itself according to the local or regional accreditation requirements. Reagents other
than those listed here may cause severe damage to the components of the instrument. Acetone,
benzene, chloroform and trichlorethane must NOT be used!