User guide

19
Leica ASP200 S
3. Instrument Components and Specifications
Nominal supply voltages: Two factory-preset voltages (not user-adjustable):
100 to 120 V or
230 to 240 V
Nominal frequency: 50 to 60 Hz
Main fuses: 2 melting fuses, 20 x 5 mm, UL-approved
• for100to120VF10A250VAC
• for230to240VF5A250VAC
Nominal power: 1000 VA
Dimensions, (L x W x H), in mm: 595 x 680 x 1325
Empty weight, unpacked: approx. 160 kg
Weight, packed: 220 kg
Operating temperature range: 15 °C to 35 °C
Relative humidity: 10 % to 80 % non-condensing
IEC 1010 classification: Protection class 1
Pollution degree 2
Overvoltage installation category II:
• 800Vimpulse(120Vsystems)
• 1500Vimpulse(240Vsystems)
Altitude: 2500 m maximum
Local/remote alarm relay: 30 V DC, maximum 2 A
2 terminals:
Each with isolated switching contact (operable both
as normally-open and normally-closed circuit)
Paraffin reservoirs
Number of containers: 3
Capacity: 4.3 l per container
Melting time: approx. 10 h
Temperature: 40 to 65 °C
Temperature accuracy: + 1 K
3.4 Technical Data