LeddarVu User Guide Page 76 of 129
Table 58: CAN bus request message (Set holding data)
Holding
Data Type
(Byte 1)
Holding Data
Description
Argument
Argument Description
0
Acquisition
configuration
Byte 2
Exponent for the number of accumulation (that
is, if the content of this register is n, 2
n
accumulations are performed).
Byte 3
Exponent for the number of oversampling (that
is, if the content of this register is n, 2
n
oversamplings are performed).
Byte 4
Number of base samples
1
Smoothing
and detection
threshold
Byte 2
Smoothing: Stabilizes the module
measurements. The behavior of the
smoothing algorithm can be adjusted by a
value ranging from 16 to 16. Clear the
Enable check box to disable smoothing.
Bytes 4
through 7
Detection threshold as a fixed-point value with
a 6-bit fractional part (that is, threshold value
is this register divided by 64).
2
Light source
power
management
Byte 2
Light source power in percentage of the
maximum. A value above 100 is an error. If a
value is specified that is not one of the
predefined values, the closest predefined
value will be used. Refer to the detection
reading to verify the LED intensity value as
seen in Table 55 and Table 67.
Note that this value is ignored if the automatic
light source intensity parameter is enabled.
Byte 3
Number of echoes for saturation acceptance:
the number of echoes can be saturated to
avoid decreasing the light source power when
using the automatic mode.
Bytes 4 and
5
Auto light source power changes the delay in
number of measurements