4.4.12. Calibration of the Static Noise
The static noise of the sensor can be calibrated by the user. This feature is
available with the integrator license available on request by contacting
LeddarTech customer service. See Section 4.4.10 to have more details about
how to add a new license to Leddar
TM
Configurator.
NOTE: When the calibration is completed, the user cannot roll back to the
factory calibration.
To perform the calibration, a shield must be beforehand installed on the
reception optics to block scattered light as it is shown in the following figure:
Figure 30: LeddarOne
TM
with Light Shielding Installed on its Lens
If the light shield is not correctly installed, a warning message will be
prompted to the user before the execution of the calibration. This message
can be ignored to continue the calibration process. The message should be
ignored only if the user wants to improve the accuracy when the sensor is
used inside a custom enclosure. For instance, a glass located in front of the
sensor can create light reflections degrading the sensor accuracy. The impact
of these reflections can be attenuated using the static noise removal
algorithm. Thereafter, under Device Calibrate, click on “Calibrate…