User manual
Serial Data Debug Solutions 
182 
919586 RevA 
PLEASE NOTE THE FOLLOWING:  
  Some additional reserved and vendor-specific FIS values are also available for selection. 
  Specific FIS values may require additional information. Consult the Serial ATA specification for specific FIS 
type details at www.sata-io.org/. 
Data Payload Pattern Type Selection Controls 
Remaining controls on the dialog vary based on a Data Payload Pattern Type selection in the following manner: 
Now, provide values for the controls on the dialog as desired. 
  Select an Operator from Equal = or Unequal <> values. Selecting Unequal triggers on all data patterns not 
matching the one specified. 
Use the remaining controls to specify your desired data pattern. The byte order follows a little endian 
representation where the least-significant byte is ordered first. Within each byte, the bits must be arranged in a 
most-significant bit first order (msb). 
  Provide a value in the At Position (in Dword) control.  
  Enter a bit string in the Data (1 - 128 bytes) control. 
Protocol Error Pattern Type Selection Controls 
Remaining controls on the dialog vary based on a Protocol Error Pattern Type selection in the following manner: 
Mark the checkbox adjacent to the protocol error type on which you wish to trigger. Protocol error types 
include: Symbol Violation, Disparity Error, Align Error, Frame Type Error, Frame Length Error, and CRC Error. 
Bus Condition Pattern Type Selection Controls 
Remaining controls on the dialog vary based on a Bus Condition Pattern Type selection in the following manner: 
Mark the checkbox adjacent to the bus condition on which you wish to trigger. Bus conditions include: Electrical 
Idle On, Electrical Idle Off, Electrical Burst, COMINIT/COMRESET, and COMWAKE. 
Note: The Trigger dialog's tab label indicates your Setup Mode selection for convenience. 