Note: The Cycle Count fields are shown, but disabled when TSS (Start) is selected as the FlexRaybus 
trigger. 
  Condition - Select from Equal, Not Equal, Less than, Less than or Equal to, Greather than, Greater than 
or Equal to, In Range, or Out Range conditions available. The default setting is Equal. 
  Value - Use this field's keypad to enter the desired Frame ID. 
  To - When the condition is set to In Range of or Out of Range, select a To value specifying the full ID 
range for the trigger. 
  Repetition Factor - When the condition is set to Equal, this field can be set to a value of 1, 2, 4, 8, 16, 32 
or 64 for triggering when Cycle multiplexing is used. 
7. Frame Qualifiers - Frame Trigger Setup Only 
Defined in the FlexRay specification, these fields allow an additional level of complexity in creating a very 
powerful FlexRay trigger. The default Qualifier setting is Don’t Care, each field can be set to One, Zero or 
Don’t Care as independent variables in the trigger setup. 
Note: The Frame Qualifiers fields are shown, but disabled when TSS (Start) is selected as the FlexRaybus 
trigger. 
SYMBOL TRIGGER SETUP DETAIL 
4b. Trigger On - Symbol Trigger Setup Only 
When Symbol is selected as the FlexRaybus trigger, use the Trigger On fields as follows: 