User manual
Operator's Manual 
919586 RevA 
33 
All TD packages provide basic tools to characterize embedded controller performance. The tools can be used on 
the decoded channels, memories, zooms, functions, etc. just like they are used on any un-decoded channels, 
memories, zooms, functions, etc. You also can use normal Edge or SMART Triggers on an analog channel input to 
trigger the oscilloscope when a certain analog signal occurs, and then measure to a particular serial data 
message using the decoded info as your guide. 
In general, some of these standard tools require a fair amount of manual setup. If the goal is to make many 
hundreds or thousands of measurements, you might want to consider using functionality built into the 
PROTObus MAG toolset. 
Note: The following examples use CANbus messages; however, similar needs exist for other serial data signals, 
and the included oscilloscope tools described in the following sections can be applied in the same way. 
Take the following example of an analog signal creating a burst of CAN messages: 
This data was acquired over a 500 ms duration. It is likely you want to understand whether the analog signal 
input to your electronic control unit (ECU) is creating the desired CAN message output from the ECU. There are a 
few ways this can be done as the following topics explain. 