User manual
Serial Data Debug Solutions 
142 
919586 RevA 
BRIEF HS COMMON MODE TEST DESCRIPTIONS 
  With HS Common Mode selected in the Category field, the following Measure values are available: 
VCMTX-DATA-1, VCMTX-DATA-0, VCMTX-CLK-1, VCMTX-CLK-0, dVCMTXLF-DATA, dVCMTXHF-DATA, 
dVCMTXLF-CLK, and dVCMTXHF-CLK. 
Name 
Purpose 
VCMTX-DATA-
1 
To verify that the Static Common-Mode Voltages (VCMTX(1), and VCMTX(0)) of the DUT Data & 
Clock Lanes HS transmitter are within the conformance limits. 
VCMTX-DATA-
0 
VCMTX-CLK-1 
VCMTX-CLK-0 
dVCMTXLF-
DATA 
To verify that the AC Common-Mode Signal Level Variations between 50 and 450MHz 
(ΔVCMTX(LF)) of the DUT Data & Clock Lanes HS transmitter are less than the maximum 
allowable limit. 
dVCMTXLF-
CLK 
dVCMTXHF-
DATA 
To verify that the AC Common-Mode Signal Level Variations above 450MHz (ΔVCMTX(HF)) of the 
DUT Data & Clock Lanes HS transmitter are less than the maximum allowable limit. 
dVCMTXHF-
CLK 
BRIEF HS LEVELS TEST DESCRIPTIONS 
  With HS Levels selected in the Category field, the following Measure values are available: 
VOD-DATA-1, VOD-DATA-0, VOD-CLK-1, VOD-CLK-0, VOHHS-Dp, VOHHS-Dn, VOHHS-CLKp, and VOHHS-
CLKn. 
Name 
Purpose 
VOD-DATA-1 
To verify that the Differential Voltages (VOD(0) and VOD(1)) of the DUT Data & Clock 
Lanes HS transmitter are within the conformance limits. 
VOD-DATA-0 
VOD-CLK-1 
VOD-CLK-0 
VOHHS-Dp 
To verify that the Single-Ended Output High Voltages (VOHHS(DP) and VOHHS(DN)) of the 
DUT Data & Clock Lanes HS transmitter are less than the maximum conformance limit. 
VOHHS-Dn 
VOHHS-CLKp 
VOHHS-CLKn 