User manual
CANbus Trigger, Decode, and Measure 
64 
CANbus-TD-TDM-OM-E RevB 
Creating a Trigger Condition 
The CAN Trigger dialog, with detail on some of the setup conditions, is shown in the images that 
follow:  
Selection of Frame Type, ID Condition, and Data Condition results in dynamic changes to the 
CAN Trigger dialog.  However, for simplicity’s sake, we will not describe all the possible 
combinations in this manual. 
To select a value for any of the conditions, touch the existing value (using your finger, or use a 
mouse pointer) to open the pop-up dialog box with a list of choices, and select one of the 
choices. 
Trigger Setup Detail 
Bit Rate -- Adjust the bit rate value here to match the bit rate 
on the bus you are connected to. This bit rate selection is 
dynamically linked to the decoding bit rate (they are always the 
same value). Use the arrows to move through standard bit 
rates (10, 25, 33.333, 50, 83.333, 100, 125, 250, 500, and 1000 
kb/s) and make a selection. Or, touch the number twice (with a 
finger, or using a mouse) and open a pop-up keypad and enter 
the value directly.  Note that if you select a value that is higher 
than that supported by the installed Trigger Coupler, you will 
not be able to trigger, and will create error frames on your bus. 
Non-standard bit rates (such as 102.4 kb/s) can be entered by 