User manual
CANbus Trigger, Decode, and Measure 
22 
CANbus-TD-TDM-OM-E RevB 
Setting Up CANbus Parameters using the Measure Setup option 
You can also set up measurement parameters using the Measure menu as follows: 
Touch Measure→Measure Setup. select the appropriate CAN or other parameter. For more 
information, see Parameter Setup in the oscilloscope's Online Help or Operator's Manual. 
Note: All CAN measurement parameters (with the exception of CAN Load %) calculate as many 
values as possible during each acquisition. If there are 10 timing events that meet the set 
condition during a specific acquisition, 10 timing event measurements will be returned. 
However, the VALUE shown in the measurement table is the last measurement made. To view 
statistical data (i.e., number of measurements made, mean, min, max, std. dev.), turn Statistics 
ON. 
CAN Load % returns only one value during each acquisition, since it is evaluating the load % for 
the entire acquisition time. 
CAN-to-Analog or Analog-to-CAN Measurement Parameter 
CAN Message to Analog Signal timing (CANtoAnalog) 
This measurement parameter is used to measure timing from either a CAN Message to an 
Analog Signal, or from an Analog Signal to a CAN Message. The Frame Type, ID, DATA, etc. 
conditions for the CAN message can be fully defined, as can the slope, level, etc. conditions for 
the analog signal transition. 
If measuring from CAN to Analog, the timing is always measured from the End of Frame of CAN 
message to the analog signal transition. If measuring from Analog to CAN, the timing is always 
measured from the analog signal transition to the Start of Frame of the CAN message. 