Aux Output Settings
The Aux Out port on the back of the oscilloscope outputs a 3.3V TTL pulse to another device following a
trigger event.
Use the Aux Output dialog to configure the calibration signal output from the Cal Out hook on the front of the
oscilloscope. The calibration output voltage is 3V.
To turn on the calibration signal:
1. Under Use Calibration Output For, choose Square wave.
2. Either enter the Frequency, or use the Set to 1 KHz Square Wave button.
Touch Off to disable the calibration signal.
Date/Time Settings
Date/Time settings control the oscilloscope's timestamp. These numbers appear in the oscilloscope
message bar and on tables/records internal to the oscilloscope application, such as History Mode and
WaveScan.
NOTE: This is not the same as the Timebase reference clock used to synchronize traces.
To access the Date/Time dialog, choose Utilities > Utilities Setup from the menu bar, then touch the
Date/Time tab.
Manual Method
Enter the Hour, Minute, Second, Day, Month, and Year, then touch the Validate Changes button.