User manual

Operator's Manual
Certifications
EMC Compliance
EC Declaration of Conformity- EMC
The oscilloscope meets intent of EC Directive 2004/108/EC for Electromagnetic Compatibility. Compliance
was demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 61326-1:2013, EN 61326-2-1:2013 EMC requirements for electrical equipment for measurement, control,
and laboratory use.
1
ELECTROMAGNETIC EMISSIONS:
EN 55011:2010, Radiated and Conducted Emissions Group 1, Class A
2 3
EN 61000-3-2/A2:2009 Harmonic Current Emissions, Class A
EN 61000-3-3:2008 Voltage Fluctuations and Flickers, Pst = 1
ELECTROMAGNETIC IMMUNITY:
EN 61000-4-2:2009 Electrostatic Discharge, 4 kV contact, 8 kV air, 4 kV vertical/horizontal coupling planes
4
EN 61000-4-3/A2:2010 RF Radiated Electromagnetic Field, 3 V/m, 80-1000 MHz; 3 V/m, 1400 MHz - 2 GHz; 1
V/m, 2 GHz - 2.7 GHz
EN 61000-4-4/A1:2010 Electrical Fast Transient/Burst, 1 kV on power supply lines, 0.5 kV on I/O signal data
and control lines
4
EN 61000-4-5:2006 Power line Surge, 1 kV AC Mains, L-N, L-PE, N-PE
4
EN 61000-4-6:2009 RF Conducted Electromagnetic Field, 3 Vrms, 0.15 MHz - 80 MHz
EN 61000-4-11:2004 Mains Dips and Interruptions, 0%/1 cycle, 70%/25 cycles, 0%/250 cycles
4 5
1 To ensure compliance with all applicable EMC standards, high quality shielded interface cables should be used.
2 Emissions which exceed the levels required by this standard may occur when the oscilloscope is connected to a test object.
3 This product is intended for use in nonresidential areas only. Use in residential areas may cause electromagnetic interference.
4 Meets Performance Criteria “B” limits of the respective standard: during the disturbance, product undergoes a temporary
degradation or loss of function or performance which is self-recoverable.
5 Performance Criteria “C” applied for 70%/25 cycle voltage dips and for 0%/250 cycle voltage interruption test levels
per EN61000-4-11.
117