Operator's Manual WaveSurfer 3000 Oscilloscopes
WaveSurfer 3000 Oscilloscopes Operator's Manual © 2014 Teledyne LeCroy, Inc. All rights reserved. Unauthorized duplication of Teledyne LeCroy documentation materials other than for internal sales and distribution purposes is strictly prohibited. However, clients are encouraged to distribute and duplicate Teledyne LeCroy documentation for their own internal educational purposes. WaveSurfer and Teledyne LeCroy are trademarks of Teledyne LeCroy, Inc.
Operator's Manual Contents Safety Instructions Symbols Precautions Operating Environment Cooling Power 1 1 1 2 2 2 Start Up Setting Up the Oscilloscope Powering On/Off Software Activation 4 4 5 6 Inputs/Outputs Front Input/Output Panel Back Input/Output Panel Analog Inputs Probes Digital Inputs 7 7 7 8 8 8 Touch Screen Menu Bar Signal Display Grid Descriptor Boxes Dialogs Turning On/Off Traces Annotating Traces Entering/Selecting Data Print Preview Printing/Screen Capture Language Selection 10 10 11
WaveSurfer 3000 Oscilloscopes Probe Settings Auto Setup Restore Default Setup Viewing Status 27 28 28 29 Digital (Mixed Signal) Digital Traces Digital Group Set Up Digital Display Set Up Renaming Digital Lines 30 30 30 31 32 Timebase Timebase Settings Sampling Modes History Mode 33 33 34 38 Trigger Trigger Modes Trigger Types Setting Up Triggers Trigger Holdoff 40 40 41 42 58 Display Display Settings Persistence 60 60 61 Cursors Cursor Types Cursor Settings 63 63 64 Measure Setting Up Measuremen
Operator's Manual Remote Control Settings Hardcopy Settings Aux Output Settings Date/Time Settings Options Preferences Settings Calibration Settings Acquisition Settings E-Mail Miscellaneous Settings Digital Voltmeter WaveSource Automatic Waveform Generator 87 90 92 92 93 93 94 95 95 96 97 99 Save/Recall Save/Recall Setups Save/Recall Waveforms Save Table Data Auto Save Disk Utilities 101 101 103 106 107 107 LabNotebook Create Notebook Entry Print to Notebook Entry Flashback Recall Configure LabNotebook
WaveSurfer 3000 Oscilloscopes Welcome Thank you for purchasing a Teledyne LeCroy WaveSurfer Oscilloscope. We're certain you'll be pleased with the detailed features unique to our instruments. The manual is arranged in the following manner: l l Safety contains important precautions and information relating to power and cooling. The sections from Start Up through Maintenance cover everything you need to know about the operation and care of the oscilloscope.
Operator's Manual Safety Instructions Observe these instructions to keep the instrument operating in a correct and safe condition. You are required to follow generally accepted safety procedures in addition to the precautions specified in this section. The overall safety of any system incorporating this instrument is the responsibility of the assembler of the system.
WaveSurfer 3000 Oscilloscopes Do not operate with suspected failures. Do not use the product if any part is damaged. Obviously incorrect measurement behaviors (such as failure to calibrate) might indicate impairment due to hazardous live electrical quantities. Cease operation immediately and sequester the instrument from inadvertent use. Operating Environment Temperature: 0 to 50° C.
Operator's Manual Power Consumption Maximum power consumption with all accessories installed (e.g., active probes, USB peripherals, digital leadset) is 150 W (150 VA) for four-channel models and 100 W (100 VA) for two-channel models. Power consumption in standby mode is 4 W. Ground The AC inlet ground is connected directly to the frame of the instrument. For adequate protection again electric shock, connect to a mating outlet with a safety ground contact. WARNING.
WaveSurfer 3000 Oscilloscopes Start Up Setting Up the Oscilloscope Carrying and Placing the Oscilloscope The oscilloscope’s case contains a built-in carrying handle. Grasp the handle firmly and lift the instrument. Always unplug the instrument from the power source before lifting and carrying it. Place the instrument where it will have a minimum 15 cm (6 inch) clearance from the nearest object. Be sure there are no papers or other debris beneath the oscilloscope or blocking the cooling vents. CAUTION.
Operator's Manual EXTERNAL MONITOR WaveSurfer 3000 supports external monitors with 1024 x 600 ppi resolution. Connect the monitor cable to the VGA video output on the back of the instrument. The connection is “plug-and-play” and does not require any further configuration on the oscilloscope. If necessary, configure the monitor to receive output. PRINTER WaveSurfer 3000 supports PictBridge-compliant printers. Connect the printer to any host USB port. The connection is "plug-and-play.
WaveSurfer 3000 Oscilloscopes Software Activation The oscilloscope operating software (firmware and standard applications) is active upon delivery. At powerup, the oscilloscope loads the software automatically. Firmware Free firmware updates are available periodically from the Teledyne LeCroy website at: teledynelecroy.com/support/softwaredownload. Registered users can receive an email notification when a new update is released. Follow the instructions on the website to download and install the software.
Operator's Manual Inputs/Outputs Front Input/Output Panel A. Power button. B. Channel inputs 1-4 for analog signals. C. Front-mounted host USB ports for transferring data or connecting peripherals such as a mouse or keyboard. D. Ground and calibration output terminal used to compensate passive probes. Back Input/Output Panel A. WaveSource connector outputs signal generated by the internal waveform generator. B. MicroSD Card slot. C. EXT Trig connector accepts external trigger. D.
WaveSurfer 3000 Oscilloscopes Analog Inputs A series of BNC connectors arranged on the front of the oscilloscope are used to input analog signal on Channels 1-4. EXT, on the back of the oscilloscope, can be used to input an external trigger pulse. Channel connectors use the ProBus interface. The ProBus interface contains a 6-pin power and communication connection and a BNC signal connection to the probe. It includes sense rings for detecting passive probes and accepts a BNC cable connected directly to it.
Operator's Manual Grounding Leads Each flying lead has a signal and a ground connection. A variety of ground extenders and flying ground leads are available for different probing needs. To achieve optimal signal integrity, connect the ground at the tip of the flying lead for each input used in your measurements. Use either the provided ground extenders or ground flying leads to make the ground connection.
WaveSurfer 3000 Oscilloscopes Touch Screen The touch screen is the principal viewing and control center of the oscilloscope. The entire display area is active: use your finger or a stylus to touch, touch-and-drag, or draw a selection box. Many controls that display information also work as “buttons” to access other functions. If you have a mouse installed, you can click anywhere you can touch to activate a control; in fact, you can alternate between clicking and touching, whichever is convenient for you.
Operator's Manual Signal Display Grid The grid area displays the waveform traces. It is sectioned into 10 Horizontal (Time) divisions and 8 Vertical (Voltage) divisions. By default, the oscilloscope divides the screen into a maximum of three grids, one each for channels/memories, math functions, and zooms. All traces of the same type appear on the same grid.
WaveSurfer 3000 Oscilloscopes Pre/Post-trigger Delay - A small arrow to the bottom left or right of the grid indicates that a pre- or post-trigger Delay has shifted the Trigger Position indicator to a point in time not displayed on the grid. All trigger Delay values are shown on the Timebase Descriptor Box. Trigger Level - This small triangle at the right edge of the grid tracks the trigger voltage level.
Operator's Manual Channel Descriptor Box Channel trace descriptor boxes correspond to analog signal inputs. They show Vertical settings and any Vertical cursor readouts: (clockwise from top left) Trace Number (Cx), Pre-Processing List (summarizes changes from default state), Coupling, Gain Setting, Offset Setting, and Averaging Sweeps Count. Codes are used to indicate pre-processing that has been applied to the input. The codes have a long and short form.
WaveSurfer 3000 Oscilloscopes Dialogs Dialogs appear at the bottom of the display for entering setup data. The top dialog will be the main entry point for the selected setup option. For convenience, related dialogs appear as a series of tabs behind the main dialog. Touch the tab to open the dialog. Right-Hand Dialogs At times, your selections will require more settings than normally appear (or can fit) on a dialog, or the task commonly invites further action, such as zooming a new trace.
Operator's Manual Turning On/Off Traces Analog Trace From the display, choose Vertical > Channel <#> Setup to turn on the trace. To turn it off, clear the Trace On checkbox on the corresponding Channel dialog. From the front panel, press the Channel button (1-4) to turn on the trace; press again to turn it off. Digital Trace From the display, choose Vertical > Digital <#> Setup. From the front panel, press the Dig button, then check Group on the Digital<#> trace dialog. Clear Group to turn off the trace.
WaveSurfer 3000 Oscilloscopes Annotating Traces The Label function gives you the ability to add custom annotations to traces that are shown on the display. Labels are numbered sequentially in the order they were created. Once placed, labels can be moved to new positions or turned off. Create Label 1. Touch the trace and choose Set label... from the context menu, or touch the trace descriptor box twice and touch the Label toolbar button on the setup dialog. 2.
Operator's Manual Entering/Selecting Data Touch Touch once to activate a control. In some cases, you’ll immediately see a pop-up menu of options. Touch one to select it. TIP: You can touch the Icon or List buttons where they appear on larger pop-ups to change how menu options are displayed. In other cases, data entry fields appear highlighted in blue when you touch them. When a data entry field is highlighted, it is active and can be modified by using the front panel Adjust knob.
WaveSurfer 3000 Oscilloscopes Touch & Drag Touch-and-drag cursor lines and annotation labels to reposition them on the grid; this is the same as setting the values on the dialog. Touch-and-drag to draw a selection box around part of a trace to quickly zoom that portion.
Operator's Manual Print Preview The Print Preview feature allows you to pause the fast display for closer waveform inspection or printing. No function other than printing is available when in Print Preview mode. All oscilloscope analysis functions (such as measurement calculations) are also paused. There are three ways to invoke Print Preview: l l l On the Utilities > Utilities Setup > Hardcopy dialog, check the box labeled Preview on Print.
WaveSurfer 3000 Oscilloscopes Printing/Screen Capture The Print function captures an image of the display and outputs it according to your Hardcopy settings. There are three ways to print a capture of the screen: l Touch the front panel Print button. l Choose File > Print. l Choose Utilities > Utilities Setup > Hardcopy tab and touch the Print button to the far right of the dialog.
Operator's Manual Front Panel Most front panel controls duplicate functionality available through the touch screen display and are described on the following pages. All the knobs on the front panel function one way if turned and another if pushed like a button. The top label describes the knob’s “turn” action, the bottom label its “push” action. Front panel buttons light up to indicate which traces and functions are active. Actions performed from the front panel always apply to the active trace.
WaveSurfer 3000 Oscilloscopes Horizontal Controls The Delay knob changes the Trigger Delay value (S) when turned. Push the knob to reset Delay to zero. The Horizontal Adjust knob sets the Time/division (S) of the oscilloscope acquisition system when the trace source is an input channel. The Time/div value is shown on the Timebase descriptor box. When using this control, the oscilloscope allocates memory as needed to maintain the highest sample rate possible for the timebase setting.
Operator's Manual Adjust and Intensity Controls The Adjust knob changes the value in any highlighted data entry field when turned. Pushing the Adjust knob toggles between coarse (large increment) or fine (small increment) adjustments when the knob is turned. When more data is available than can actually be displayed, the Intensity button helps to visualize significant events by applying an algorithm that dims less frequently occurring samples.
WaveSurfer 3000 Oscilloscopes Zooming Waveforms The Zoom function magnifies a selected region of a trace. On WaveSurfer 3000 model oscilloscopes, you can display up to four zoom traces (Z1-Z4) taken from any channel, math, or memory trace. Creating Zooms To create a zoom, touch -and-drag to draw a selection box around any part of the source waveform. Selection box over trace. The zoom will resize the selected portion to fit the full width of the grid.
Operator's Manual and memory traces). Turn off Zoom To close the zoom, either touch the zoom descriptor box twice to open the Zoom dialog and deselect Trace On, or touch the zoom trace to open the context menu and choose Off. Zoom Controls To open the Zoom dialog, touch twice on any zoom descriptor box, or choose Math > Zoom Setup from the menu bar. Trace Controls Trace On shows/hides the zoom trace. It is selected by default when the zoom is created.
WaveSurfer 3000 Oscilloscopes Vertical Vertical, also called Channel, settings usually relate to voltage level and control the trace along the Y axis. NOTE: While Digital settings can be accessed through the Vertical menu on oscilloscopes with the Mixed Signal option, they are handled quite differently. See Digital. The amount of voltage displayed by one vertical division of the grid, or Vertical Scale (V/div), is most quickly adjusted by using the front panel Vertical knob.
Operator's Manual Probe Settings When a Teledyne LeCroy-compatible probe is connected to the oscilloscope input, the probe is automatically identified and the model name displayed on the Channel dialog under the "Probe" heading. Also, the Probe dialog bearing the probe name is added to the right of the Channel dialog. When a probe is not connected, the Channel dialog shows only the Cx tab for vertical setup.
WaveSurfer 3000 Oscilloscopes DeGauss Probe The Degauss control is activated for some types of probes (e.g., current probes). Degaussing eliminates residual magnetization from the probe core caused by external magnetic fields or by excessive input. It is recommended to always degauss probes prior to taking a measurement. CAUTION. Remove the probe from the circuit under test before initializing DeGauss.
Operator's Manual Viewing Status All oscilloscope settings can be viewed through the various Status dialogs. These show all existing acquisition, trigger, channel, math function, measurement and parameter configurations, as well as which are currently active. Access the Status dialogs by choosing the Status option from the Vertical, Timebase, Trigger, or Math menus (e.g., Channel Status, Acquisition Status).
WaveSurfer 3000 Oscilloscopes Digital (Mixed Signal) The digital leadset (delivered with the WS3K-MSO option) inputs up-to-16 lines of digital data. Leads are organized into two banks of eight leads each, and you assign each bank a standard Logic Family or a custom Threshold to define the digital logic of the signal. The Digital set up dialog has two tabs each corresponding to one of two possible digital groups, labeled Digital1 to Digital2, which correspond to buses.
Operator's Manual 2. On the Digital<#> set up dialog, check the boxes for lines D0 through D15 that comprise the group. Touch the Display D0-D7 and Display D8-D15 buttons to quickly turn on the entire digital bank, or touch the Right and Left Arrow buttons to switch between each digital bank as you make line selections. NOTE: Each group can consist of anywhere from 1 to 16 of the leads that are (or will be) connected to signal, from either digital bank regardless of the Logic set on the bank.
WaveSurfer 3000 Oscilloscopes Individual traces are resized to fit the total number of divisions available. The example above shows a group of three Line traces plus the Bus trace occupying a Group Height of 4.0 divisions. Each trace takes up one division. 5. Check the Group box to enable the display. To close traces, uncheck the Group box. Renaming Digital Lines The labels used to name each line can be changed to make the user interface more intuitive. Also, labels can be "swapped" between lines.
Operator's Manual Timebase Timebase, also known as Horizontal, settings control the trace along the X axis. The timebase is shared by all channels. The time represented by each horizontal division of the grid, or Time/Division, is most easily adjusted using the front panel Horizontal knob. Full Timebase set up, including sampling mode selection, is done on the Timebase dialog, which can be accessed by either choosing Timebase > Horizontal Setup from the menu bar, or touching the Timebase descriptor box.
WaveSurfer 3000 Oscilloscopes Sampling Modes Real Time Sampling Mode Real Time sampling mode is a series of digitized voltage values sampled on the input signal at a uniform rate. These samples are displayed as a series of measured data values associated with a single trigger event. By default, the waveform is horizontally positioned so that the trigger event is time zero on the grid.
Operator's Manual SET UP SEQUENCE MODE When setting up Sequence Mode, you define the number of fixed-size segments acquired in single-shot mode (see the instrument specifications for the limits). The oscilloscope uses the sequence timebase setting to determine the capture duration of each segment. Along with this setting, the oscilloscope uses the number of segments, maximum segment length, and total available memory to determine the actual number of samples or segments, and time or points. 1.
WaveSurfer 3000 Oscilloscopes Use the Zoom controls to change the scale factors of the trace. Roll Mode Roll mode displays, in real time, incoming points in single-shot acquisitions that appear to "roll" continuously across the screen from right to left until a trigger event is detected and the acquisition is complete. The parameters or math functions connected to each channel are updated every time the roll mode buffer is updated, as if new data is available.
Operator's Manual Average Sampling Mode Average sampling mode calculates the average value for each captured point over a specified number of acquisitions (2, 4, 8, 16, 32, 64, 128, 256, 512, and 1024 sweeps are all available). Each individual acquisition uses Real Time mode and the results are averaged together. Average mode can be used to reduce random noise in repeating signals. When selecting Average sampling mode, also select the number of Sweeps to calculate in the Average.
WaveSurfer 3000 Oscilloscopes History Mode History Mode allows you to review any acquisition saved in the oscilloscope's history buffer, which automatically stores all acquisition records until full. Not only can individual acquisitions be restored to the grid, you can "scroll" backward and forward through the history at varying speeds to capture individual details or changes in the waveforms over time.
Operator's Manual l l The top row of buttons scrolls continuously and are (left to right): Fast Backward, Slow Backward, Pause, Slow Forward, Fast Forward. The bottom row of buttons steps one record at a time and are (left to right): Back to Start, Back One, Go to Index (#), Forward One, Forward to End. 5. Entering History Mode automatically stops new acquisitions. To leave History Mode, press the History Mode button again, or restart acquisition by pressing one of the front panel Trigger Mode buttons.
WaveSurfer 3000 Oscilloscopes Trigger While the oscilloscope is continuously sampling signal when it is turned on, it can only display up to its maximum memory in data samples. Triggers select an exact event/time in the waveform to display on the oscilloscope screen so that memory is not wasted on insignificant periods of the signal.
Operator's Manual Trigger Types These are the trigger types available for selection. If the trigger is part of a subgroup (e.g., Smart), first choose the subgroup from among the basic types to display all the trigger options. Basic Triggers Edge triggers upon a achieving a certain voltage level in the positive or negative slope of the waveform. Width triggers upon finding a positive- or negative-going pulse width when measured at the specified voltage level.
WaveSurfer 3000 Oscilloscopes SlewRate triggers when the rising or falling edge of a pulse crosses an upper and a lower level. The pulse edge must cross the thresholds faster or slower than a selected period of time. Serial Triggers The Serial trigger type will appear if you have installed protocol-specific serial data trigger and decode options. Select this type to open the serial trigger setup dialogs. Instructions for using all serial data options are available from our website at teledynelecroy.
Operator's Manual 3. Choose the Slope (edge) of the wave on which to trigger. 4. Choose the type of signal Coupling at the input. Choices are: l l l l DC - All the signal’s frequency components are coupled to the trigger circuit for high frequency bursts or where the use of AC coupling would shift the effective trigger level. AC - The signal is capacitively coupled. DC levels are rejected, and frequencies below 10 Hz are attenuated.
WaveSurfer 3000 Oscilloscopes Width Trigger Width triggers upon finding a positive- or negative-going pulse width when measured at the specified voltage level. On the Trigger dialog, select Width trigger type to display the controls. ANALOG WIDTH 1. Choose the Source input. 2. Choose the type of signal Coupling at the input.
Operator's Manual DIGITAL WIDTH 1. Choose the Source input line. 2. Choose the line Polarity at which to measure pulse width. 3. Choose the Logic Family that marks the High-Low logic threshold. To enter a custom threshold, choose Logic Family User Defined and enter the voltage Level. NOTE: The Logic Family will default to any Logic Setup associated with that line in a previous digital group setup. 4. Use Width Condition is settings to create an expression describing the triggering pulse width.
WaveSurfer 3000 Oscilloscopes Qualified Trigger Qualified arms the trigger on the A event, then fires on the B event. In Normal trigger mode, it automatically resets after the B event. The options for the B event depend on the type of A event. You may apply additional Holdoff by time or number of events. On the Trigger dialog, select Qualified trigger type to display the controls. Besides an Edge level, the arming ("A") event may be a State, any voltage measured above or below a threshold Level.
Operator's Manual Pattern Trigger Pattern is the default trigger when the digital leadset is connected to the oscilloscope, as these users generally wish to find and trigger upon digital logic patterns. However, a Pattern trigger can also be set on a user-defined pattern of High or Low voltage levels in analog channels (including the External Trigger input), or a combination of digital and analog patterns when Mixed Signal capabilities are available. On the Trigger dialog, select Pattern trigger type.
WaveSurfer 3000 Oscilloscopes 3. If you have not already set a logic threshold, open the Levels dialog and select a Logic Family for each digital bank from which you've selected lines. To set a custom logic threshold, choose Logic Family User Defined, then enter the Threshold voltage and Hysteresis. NOTE: Digital lines inherit the Logic Setup made when defining digital groups. However, you can change the logic threshold on the Levels dialog, as well.
Operator's Manual Standard Dialogs The standard dialog for setting up an analog Pattern trigger includes all the controls for setting the pattern and the voltage threshold on the same dialog. 1. Select the Boolean Operator (AND, NAND, OR, or NOR) that describes the relationship among analog inputs (e.g., C1 must be High NAND C2 must be Low). 2. For each input to be included in the trigger pattern, and select what State it must be in (High, Low, or Don't Care) compared to the threshold Level you will set.
WaveSurfer 3000 Oscilloscopes TV Trigger TV triggers on a specified line and field in standard (PAL, SECAM, NTSC, HDTV) or custom composite video signals. On the Trigger dialog, select TV trigger type to display the controls. 1. Choose the Source signal input. 2. Choose the signal TV Standard. To use a custom signal, also enter the: l Frame Rate l # of Fields per line l # of Lines l Interlace ratio 3. Choose the Line and Field upon which to trigger.
Operator's Manual Window Trigger Window triggers when a signal exits a window defined by voltage thresholds. On the Trigger dialog, select Smart trigger type, then choose Window to display the controls. 1. Choose the Source signal input. 2. Choose the type of signal Coupling at the input. Choices are: l l l l DC - All the signal’s frequency components are coupled to the trigger circuit for high frequency bursts or where the use of AC coupling would shift the effective trigger level.
WaveSurfer 3000 Oscilloscopes Interval Trigger Interval triggers upon finding a specific interval, the time (period) between two consecutive edges of the same polarity: positive to positive or negative to negative. Use the interval trigger to capture intervals that fall short of, or exceed, a specified range. On the Trigger dialog, select Smart trigger type, then Interval to display the controls. ANALOG INTERVAL 1. Choose the Source input. 2. Choose the type of signal Coupling at the input.
Operator's Manual DIGITAL INTERVAL 1. Choose the Source input line. 2. Choose the Slope (edge) from which to measure. 3. Choose the Logic Family that marks the High-Low logic threshold. To enter a custom threshold, choose Logic Family User Defined and enter the voltage Level. NOTE: The Logic Family will default to any Logic Setup associated with that line in a previous digital group setup. 4. Use Interval Condition is settings to create an expression describing the triggering interval.
WaveSurfer 3000 Oscilloscopes Dropout Trigger Dropout triggers when a signal loss is detected. The trigger is generated at the end of the timeout period following the last edge transition that meets the trigger conditions. It is used primarily in single-shot applications with a pre-trigger delay. On the Trigger dialog, select Smart trigger type, then Dropout to display the controls. ANALOG DROPOUT 1. Choose the Source signal input. 2. Choose the type of signal Coupling at the input.
Operator's Manual DIGITAL DROPOUT 1. Choose the Source digital line. 2. Choose the Slope (edge) to watch for transitions. 3. Choose the Logic Family that marks the transition threshold. To enter a custom threshold, choose Logic Family User Defined and enter the voltage Level. NOTE: The Logic Family will default to any Logic Setup associated with that line in a previous digital group setup. 4. Under Dropout Condition is...
WaveSurfer 3000 Oscilloscopes Runt Trigger Runt triggers when a pulse crosses a first threshold, but fails to cross a second threshold before re-crossing the first. Other defining conditions for this trigger are the polarity and runt interval (width). On the Trigger dialog, select Smart trigger type, then choose Runt to display the controls. 1. Choose the Source input. 2. Choose the type of signal Coupling at the input.
Operator's Manual SlewRate Trigger SlewRate triggers when the rising or falling edge of a pulse crosses an upper and a lower level. The pulse edge must cross the thresholds faster or slower than a selected period of time. On the Trigger dialog, select Smart trigger type, then Slew Rate to display the controls. 1. Choose the Source input. 2. Choose the type of signal Coupling at the input.
WaveSurfer 3000 Oscilloscopes Trigger Holdoff Holdoff is an additional condition that may be set for Edge and Pattern triggers. It can be expressed either as a period of time or an event count. Holdoff disables the trigger temporarily, even if the trigger conditions are met, until the holdoff conditions are also met. The trigger fires when the holdoff has elapsed. Use holdoff to obtain a stable trigger for repetitive, composite waveforms.
Operator's Manual In the figure below, the bold edges on the trigger source indicate that a positive slope has been selected. The broken, upward-pointing arrows indicate potential triggers, while the bold ones show where triggers actually occur after the holdoff expires. Edge trigger with holdoff by events. Holdoff Settings To access the Trigger Holdoff dialog, choose Triggers > Trigger Setup from the menu bar or press the front panel Trigger Setup button, then touch the Holdoff tab.
WaveSurfer 3000 Oscilloscopes Display Display settings affect the number and style of grids that appear on screen and some of the visual characteristics of traces, such as persistence. Display Settings To access the Display dialogs, choose Display > Display Setup or Display > Persistence Setup. Grid Select one of the grid types: Auto, the default, automatically adds or deletes grids as you open or close traces, up to the maximum number supported. Single displays a single grid shared by all traces.
Operator's Manual Persistence The Persistence feature retains waveform traces on the display for a set amount of time before allowing them to gradually "decay," similar to the display of old phosphor screen oscilloscopes. Use Persistence to accumulate on-screen points from many acquisitions to see your signal change over time. The persistence modes show the most frequent signal path in three-dimensional intensities of the same color (Analog), or graded in a spectrum of colors (Color).
WaveSurfer 3000 Oscilloscopes Other Persistence Settings Besides the different modes, you can select a Saturation level as a percentage of the maximum population. All populations above the saturation population are then assigned the highest color intensity: that is, they are saturated. At the same time, all populations below the saturation level are assigned the remaining intensities. Data populations are dynamically updated as data from new acquisitions is accumulated.
Operator's Manual Cursors Cursors are markers (lines or cross-hairs) that identify specific voltage and time values on the waveform. Use cursors to make fast, accurate measurements of specific points in the waveform. There are three, standard cursor types available. Vertical (amplitude) cursor readouts appear the descriptor box for the trace; Horizontal (time) cursor readouts appear below the Timbebase descriptor box. Horizontal and vertical cursors.
WaveSurfer 3000 Oscilloscopes Cursor Settings Display Cursors To quickly turn on/off cursors, either: l l From the menu bar, choose Cursors then select the desired cursor type from the drop-down list. On the front panel, press the Cursor Type button repeatedly to cycle through all the cursor types. Stop when the desired type is displayed. Position Cursors Te easiest way to position a cursor is to touch and drag the cursor line to a new position.
Operator's Manual Measure Measurement parameters are tools that give you access to a wide range of waveform properties. Use them to analyze many attributes of your waveform such as rise-time, rms voltage, and peak-to-peak voltage. You can create a custom set of up to six parameters drawn from all the standard measurements, as well as specialized measurements installed with optional software packages. Measurement readouts appear in a table below the grid. Readouts can be individually turned on/off.
WaveSurfer 3000 Oscilloscopes 5. Optionally, set a measurement gate by entering the Gate Start and Stop divisions or dragging the gate posts from the far left and right edges of the grid to reposition them. 6. Choose to display Statistics or Histicons on the measurement readout table. Gating Measurements By using gates, you can narrow the span of the waveform on which to perform parameter measurements, allowing you to focus on the area of greatest interest.
Operator's Manual List of Standard Measurement Parameters NOTE: There may be additional parameters available depending on the software options installed on the oscilloscope. Parameter Description Amplitude (ampl) Measures the difference between upper and lower levels in two-level signals. Differs from pkpk in that noise, overshoot, undershoot, and ringing do not affect the measurement. Amplitude is calculated by using the formula Top – Base.
WaveSurfer 3000 Oscilloscopes Parameter Description period is measured for each transition pair, with values averaged to give final result. Phase Phase difference between signal analyzed and signal used as reference. Both signals are measured from the 50% point of their rising edges. Rise 20-80% (rise2080) Duration of pulse waveform's rising transition from 20% to 80% of the amplitude averaged for all rising transitions between the measurement gates.
Operator's Manual Calculating Measurements Determining Top and Base Lines Proper determination of the top and base reference lines is fundamental for ensuring correct parameter calculations. The analysis begins by computing a histogram of the waveform data over the time interval spanned by the left and right measurement gates. For example, the histogram of a waveform transitioning in two states will contain two peaks (see figure).
WaveSurfer 3000 Oscilloscopes Where Mr is the number of leading edges found, Mf the number of trailing edges found, rising edge i crosses the x% level, the time when and the time when falling edge i crosses the x% level. Determining Time Parameters Time parameter measurements such as width, period and delay are carried out with respect to the mesial reference level, located halfway (50%) between the top and base reference lines.
Operator's Manual Math Math traces (F1-F2)) display the result of applying a mathematical operation to a source trace. The output of a math function is always another trace, whereas the output of a measurement parameter is a tabular readout of the measurement. Math can be applied to any channel (Cx), zoom (Zx), or memory (Mx) trace.
WaveSurfer 3000 Oscilloscopes 5. If the operator you've selected has any other configurable settings, you'll see a right-hand dialog of the same name as the operator. Touch the tab to open the dialog and make any further settings. These are explained on the dialog. There will also be a Zoom dialog where you can optionally adjust the math trace. This does not affect the scale of any other traces.
Operator's Manual List of Standard Math Functions NOTE: Additional math functions may be available depending on the options installed on the oscilloscope. Function Definition Absolute For every point in the waveform the distance away from zero is calculated. For values greater than zero this is the same as the value. For values less than zero, the magnitude of this value without regard to its sign is used. Average Calculates either a summed or continuous average of a selected number of sweeps.
WaveSurfer 3000 Oscilloscopes Function Definition Trend Produces a waveform composed of a series of parameter measurements in the order the measurements were taken. The vertical units are those of the source parameter, the horizontal unit is measurement number. The trend contains a single value for each measurement. Zoom Zoom of input source. Trend A trend is a waveform composed of a series of parameter measurements in the order the measurements were taken.
Operator's Manual Rescaling and Assigning Units The rescale function allows you to apply a multiplication factor (a) and additive constant (b) to your waveform: aX + b. You can do it in the unit of your choice, depending on the type of application. Set Up Rescaling 1. Follow the usual steps to set up a math function, selecting Rescale from the Functions submenu. 2. Touch the Rescale right-hand dialog tab. 3.
WaveSurfer 3000 Oscilloscopes Abbreviated Units of Measure Abbreviation Measure Abbreviation Measure (blank) No units N Newton A Ampere OHM Ohm C Coulomb PAL Pascal CYCLE Cycles PCT Percent DB Decibel POISE Poise DBC Decibel referred to carrier PPM Parts per million DBM Decibel Milliwatt RAD Radian DBV Decibel Volts DEG Degree (of arc) DBUZ Decibel Microamp MNT Minute (of arc) DEC Decade SAMPLE Sample DIV Divisions SWEEP Sweeps Event Events SEC Second (of
Operator's Manual Averaging Waveforms Summed Averaging Summed Averaging is the repeated addition, with equal weight, of successive source waveform records. If a stable trigger is available, the resulting average has a random noise component lower than that of a singleshot record. Whenever the maximum number of sweeps is reached, the averaging process stops. In Summed averaging, you specify the number of acquisitions to be averaged. The averaged data is updated at regular intervals.
WaveSurfer 3000 Oscilloscopes 6th sweep: new average = (new data + 4 * old average)/(4 + 1) = 1/5 new data + 4/5 old average 7th sweep: new average = (new data + 4 * old average)/(4 + 1) = 1/5 new data + 4/5 old average In this way, for sweeps > 4 the importance of the old average begins to decrease exponentially. NOTE: The number of sweeps used to compute the average is displayed at the bottom of the trace descriptor box.
Operator's Manual 3. Choose an Output type. 4. Optionally, choose a weighting Window. See below for more information about FFT weighting windows. 5. Depending on your Output Type selection, you may also make selections for : l l Group Delay Shift Line Impedence. By default, the FFT function assumes that the oscilloscope is terminated in 50 Ohms. If an external terminator is being used, this setting can be changed to properly calculate the FFT based on the new termination value. 6.
WaveSurfer 3000 Oscilloscopes Window Type Applications and Limitations Rectangular These are normally used when the signal is transient (completely contained in the time-domain window) or known to have a fundamental frequency component that is an integer multiple of the fundamental frequency of the window. Signals other than these types will show varying amounts of spectral leakage and scallop loss, which can be corrected by selecting another type of window.
Operator's Manual Memory The oscilloscope is equipped with twointernal memory slots (Mx) to which you can copy any channel, math, zoom or other special waveform that is active on the grid. This is a convenient way to store an acquisition for later viewing and analysis. Memories are created at the same scale as the source trace, but they can be adjusted independent of the original by using the Zoom controls that appear next to the Mx dialogs. Save Waveform to Memory 1.
WaveSurfer 3000 Oscilloscopes Analysis Most Teledyne LeCroy oscilloscopes calculate measurements for all instances in the acquisition, enabling you to rapidly and thoroughly analyze a long memory acquisition of thousands or millions of parameter values, or to apply a variety of mathematical functions to the waveform trace.
Operator's Manual Scan Modes The scan mode determines the type of search to be performed. Select the Mode along with the Source trace to be searched on the main WaveScan dialog. For each mode, different controls appear on the WaveScan dialog, providing additional inputs to the search criteria. Make the appropriate entries in these fields before starting the search. EDGE MODE Edge Mode is used for detecting the occurrence of edges. Events that meet the threshold level are captured and tabulated.
WaveSurfer 3000 Oscilloscopes l Runt Type. Choose Both, Pos, or Neg. l Hysteresis. Enter the hysteresis level as a percentage or voltage. l Low Threshold and High Threshold. Enter the levels as a percentage or voltage. l Absolute Levels. Check this box to enter levels as absolute voltage instead of percentage. MEASUREMENT MODE Measurement Mode is used for applying filters to measurements to find those that meet your defined criteria, helping to isolate particular events within many samples.
Operator's Manual A unique feature of the WaveScan Zoom is that you can automatically zoom the events captured from the source trace by touching the Prev/Next buttons on the Zx dialog. You can also select the event from the Table display, and you are automatically relocated to that event on the zoom trace. Setting Up WaveScan This procedure explains how to set up WaveScan to search an acquisition for events of interest. Set up your source channel and triggers before setting up the scan. 1.
WaveSurfer 3000 Oscilloscopes Utilities Utilities settings primarily control the instrument's interaction with other devices/systems. Preferences, on the other hand, tend to control the appearance and performance of the oscilloscope application. To access the Utilities dialog, choose Utilities > Utilities Setup... from the menu bar. HardCopy Setup, Date/Time Setup, and System Status buttons open their corresponding dialogs, as do the tabs.
Operator's Manual Remote Control Settings The Remote dialog contains settings to configure remote control of the instrument and also network access. Supported remote control protocols are: l l TCPIP (Ethernet). If you choose this option, also install Teledyne LeCroy's VICP drivers on the controller. These are included in the VICP Passport plug-in, available free from teledynelecroy.com.
WaveSurfer 3000 Oscilloscopes 5. On the Ethernet Driver Settings dialog, choose Specify an IP Address. If assigning a Name server, leave Obtain an IP address via DHCP. 6. Touch the IP Address field, and use the keyboard to enter the address. Repeat for Subnet Mask and Default Gateway. Alternatively, touch the Name Server tab and enter the DNS server address. 7. Touch the window close boxes to return to the oscilloscope application.
Operator's Manual Set Up Remote Control NOTE: Full remote control setup requires the installation and configuration of software on the controller, as well. These steps represent only what is done on the oscilloscope. See your oscilloscope Remote Control manual for an explantion of the complete process. 1. From the menu bar, choose Utilities > Utilities Setup, then touch the Remote tab. 2. On the Remote dialog, make a Control From selection. 3.
WaveSurfer 3000 Oscilloscopes Hardcopy Settings Hardcopy settings control how the oscilloscope Print function behaves. Print captures an image of the oscilloscope display, but there are several options as to what it does with the image next: l Send to a hardcopy printer l "Print" to a file that can be saved to an internal or external drive l Send to E-Mail Each option is set up on the Utilities Hardcopy dialog. You can further set up a default print color scheme and capture area.
Operator's Manual 4. Enter a File Name. This will form the basis of all filenames, until you change it. NOTE: Numbers you add to the end of the filename will be truncated, as the instrument appends numbers to this name with each new file. To add your own identifying numbers, place them at the front of the name. 5. To save the file on a USB drive instead of the internal storage card, touch Browse next to Directory and choose USB Disk.
WaveSurfer 3000 Oscilloscopes Aux Output Settings The Aux Out port on the back of the oscilloscope outputs a 3.3V TTL pulse to another device following a trigger event. Use the Aux Output dialog to configure the calibration signal output from the Cal Out hook on the front of the oscilloscope. The calibration output voltage is 3V. To turn on the calibration signal: 1. Under Use Calibration Output For, choose Square wave. 2. Either enter the Frequency, or use the Set to 1 KHz Square Wave button.
Operator's Manual Internet Method This method uses the Simple Network Time Protocol (SNTP) to read the time from time-a.nist.gov. The oscilloscope must be connected to an internet access device through the LAN (Ethernet) port on your instrument's I/O panel. . If your connection is active, touch the Set from Internet button. Time Zone Select a Time Zone. Check Auto Daylight Savings Time to automatically reset the timestamp when daylight savings occurs in the selected time zone.
WaveSurfer 3000 Oscilloscopes Calibration Settings To ensure the instrument maintains specified performance, it is factory set to perform a calibration during warm-up. We recommend that you warm up the oscilloscope for at least 20 minutes prior to use to give the instrument time to complete calibration procedures. Manually calibrate the oscilloscope when: l It is used in temperatures that differ from the previous calibration temperature by more than 5° C.
Operator's Manual Acquisition Settings The Acquisition settings determine how traces behave on screen as gain or timebase changes. Offset Setting constant in: l l Volts moves the vertical offset level indicator with the actual voltage level. Div(isions) keeps the vertical offset level indicator stationary. The waveform remains on the grid as you increase the gain; whereas, if Volts is selected, the waveform could move off the grid.
WaveSurfer 3000 Oscilloscopes If desired, enter a Default Recipient Address (To:) All email sent from the oscilloscope will go to this address unless manually changed. Click Send Test Mail to send a confirmation message to ensure proper e-mail configuration. Miscellaneous Settings These other Preference settings are located on the Miscellaneous dialog. To add the Teledyne LeCroy logo to print output, check Print Teledyne LeCroy Logo When Printing Grid Area Only.
Operator's Manual Digital Voltmeter The Digital Voltmeter option (WS3K-DVM) activates an integrated 4-digit digital voltmeter and 5-digit frequency counter that operates through the same probes already attached to the oscilloscope channels. Use it to view real-time measurements through a dedicated user interface display that continues even when your triggering system is stopped. NOTE: You do not need to display the source channel trace to see the voltage readout on the DVM descriptor box and dialog.
WaveSurfer 3000 Oscilloscopes Set Up DVM 1. From the menu bar, choose Utilities > DVM. 2. Mark Enable to turn on the DVM function. 3. Touch Source and select the input channel to be measured. 4. Touch Mode and choose the waveform parameter to display on the readout. Options are: l DC l DC RMS l AC RMS l Frequency 5.
Operator's Manual WaveSource Automatic Waveform Generator The WaveSource Automatic Waveform Generator allows you to output custom sine, square, triangle, pulse, DC, noise, and arbitrary waveforms from the WaveSurfer 3000. Connect a BNC cable from the WaveSource output on the back of the oscilloscope to the external device. Output is continuous once WaveSource is enabled. Output Standard Waveform 1. Choose Utilities > WaveSource or touch the front panel WaveSource button. 2. Select the Load level. 3.
WaveSurfer 3000 Oscilloscopes 1. Choose Utilities > WaveSource or touch the front panel WaveSource button. 2. Select the Load level. 3. Choose Waveform type ARB. 4. Touch Browse next to the File Name field and browse to the location of the arbitrary waveform file on either the Storage Card or the USB Disk. Select the file and touch OK. 5. Touch Upload. 6. Check Enable to begin output.
Operator's Manual Save/Recall The oscilloscope File menu allows you to save or retrieve waveform files, measurement table data, and instrument setup panels. Access these functions by choosing any of the Save or Recall options from the File menu. The dialog contains a tab for each file function. Save/Recall Setups Save Setups allows you to quickly save up-to-six oscilloscope panel settings to internal storage, while Recall Setups restores them with a touch.
WaveSurfer 3000 Oscilloscopes Recalling Oscilloscope Setups Choose File > Recall Setup... from the menu bar. RECALL SETUP FROM MEMORY Touch one of the six Recall buttons under Recall From Internal Setup.... NOTE: If a setup has been stored to a location, it is labeled with the save date/time. Otherwise, the slot is labeled Empty. RECALL SETUP FROM FILE 1. In Recall panel from file, touch Browse and navigate to the desired folder. 2. Select the setup file and touch OK. 3.
Operator's Manual Save/Recall Waveforms The Save Waveform function saves trace data to either an internal memory location, or to a text or binary format file (.trc). The source waveform can be any trace; for example, a channel, math function, or a waveform stored in another memory. Use Recall Waveform to restore these previously saved waveforms to the display. NOTE: Only .trc files saved in binary format can be recalled into the oscilloscope.
WaveSurfer 3000 Oscilloscopes 4. Touch Data Format and select a file format: l Binary, Teledyne LeCroy's binary file format (.trc). Binary results in the smallest possible file size, and is necessary for recalling waveforms to Teledyne LeCroy instruments. NOTE: Binary files can be converted to ASCII using Teledye LeCroy utilities such as ScopeExplorer or WaveStudio. l ASCII text file (.txt extension). l MATLAB text file (.dat extension). l Excel text file (.csv extension). l MathCad text file (.
Operator's Manual Recall Waveform Choose File > Recall Waveform from the menu bar. NOTE: Only .trc files saved in binary format can be recalled into the oscilloscope. RECALL WAVEFORM FROM MEMORY 1. Touch Memory. 2. Touch Source and choose a memory location from the Select Source pop-up. 3. Touch Destination and select a location into which to open the recalled memory. 4. Mark Show on Recall to display the trace on the grid. 5. Touch Recall Now! RECALL WAVEFORM FROM FILE 1. Touch File. 2.
WaveSurfer 3000 Oscilloscopes Save Table Data The Save Table function saves tabular measurement data displayed on screen to an Excel or ASCII file. By default, files are saved on the MicroSD card, although you can choose a USB drive. Access the Save Table dialog by choosing File > Save Table from the menu bar. 1. Leave the default Source selection All Displayed. 2. Optionally touch Table Title and enter a new root file name. CAUTION.
Operator's Manual Auto Save Data that appears on the oscilloscope display, such as waveforms, measurement readouts and decoder data, can be very dynamic and difficult to read from the oscilloscope unless you stop the acquisition. The Auto Save enables you to automatically store waveform and table data to a file that can be recalled to the oscilloscope later or saved permanently to external storage.
WaveSurfer 3000 Oscilloscopes Delete All Files in a Folder 1. Touch the Delete button. 2. Browse to the current folder containing the file. 3. With the desired folder selected, touch Empty Folder. Create a New Folder 1. Touch Create. 2. Touch Current folder and provide the full path to the new folder, including the folder name. 3. Touch Create Folder. Back Up Files Touch Backup to back up the entire contents of the hard drive to a removable storage device.
Operator's Manual LabNotebook Teledyne LeCroy's LabNotebook feature extends the documentation capabilities of your oscilloscope. It allows you to create and save Notebook Entries containing all displayed waveforms, the oscilloscope setup under which they were taken, and custom annotations. Notebook Entries are stored in an internal database and are available for recall at any time. Besides storing the waveform data, LabNotebook also stores your panel setups and parameter measurements.
WaveSurfer 3000 Oscilloscopes Print to Notebook Entry The front panel Print button can be configured to capture the display and create a new Notebook Entry. This is a convenient way to create new Notebook Entries as you work. To configure the Print button for Notebook Entries, go to File > LabNotebook > Preferences tab and check Create Entry when Hardcopy Pressed. Flashback Recall Once a Notebook Entry is made, you can recall it at any time using Flashback Recall.
Operator's Manual Configure LabNotebook Preferences To configure the behavior of the LabNotebook tool, on the menu bar, choose File > Lab Notebook, then touch the Preferences tab. Select/deselect the following options: Prompt for Entry Title Before Saving opens the LabNotebook dialog when a new entry is created. You can elect to name notebook entries using only the date/timestamp by leaving this box unchecked.
WaveSurfer 3000 Oscilloscopes Maintenance Cleaning Clean only the exterior of the instrument using a soft cloth moistened with water or an alcohol solution. Do not use harsh chemicals or abrasive elements. Under no circumstances submerge the instrument or allow moisture to penetrate it. Avoid electric shock by unplugging the power cord from the AC outlet before cleaning. CAUTION. Do not attempt to clean internal parts. Refer to qualified service personnel.
Operator's Manual Touch Screen Calibration Periodically calibrate the touch screen to maintain its accuracy and responsiveness. We recommend that you use a stylus rather than your finger for this procedure. 1. From the menu bar, choose Utilities > Utilities Setup. 2. On the Utilities main dialog, touch Touch-Screen Calibration. 3. Following the prompts, touch as close as possible to the center of each cross that appears on the screen until the calibration sequence is complete.
WaveSurfer 3000 Oscilloscopes WaveSurfer 3000 Firmware Update Teledyne LeCroy frequently releases free firmware updates for X-Stream model oscilloscopes containing new product features and bug fixes. The X-Stream installer updates multiple components including the oscilloscope application, required DLLs, drivers, and low-level microcode for integrated circuits on the oscilloscope.
Operator's Manual Returning a Product for Service Contact your local Teledyne LeCroy service center for calibration or other service. If the product cannot be serviced on location, the service center will give you a Return Material Authorization (RMA) code and instruct you where to ship the product. All products returned to the factory must have an RMA. Return shipments must be prepaid. Teledyne LeCroy cannot accept COD or Collect shipments. We recommend air-freighting.
WaveSurfer 3000 Oscilloscopes Contact Teledyne LeCroy United States and Canada - World Wide Corporate Office Teledyne LeCroy Corporation 700 Chestnut Ridge Road Chestnut Ridge, NY, 10977-6499, USA Ph: 800-553-2769 / 845-425-2000 FAX: 845-578-5985 teledynelecroy.com Support: contact.corp@teledynelecroy.com Sales: customersupport@teledynelecroy.com United States Protocol Solutions Group Teledyne LeCroy Corporation 3385 Scott Boulevard Santa Clara, CA, 95054, USA FAX: 408-727-0800 teledynelecroy.
Operator's Manual Certifications EMC Compliance EC Declaration of Conformity- EMC The oscilloscope meets intent of EC Directive 2004/108/EC for Electromagnetic Compatibility. Compliance was demonstrated to the following specifications as listed in the Official Journal of the European Communities: EN 61326-1:2013, EN 61326-2-1:2013 EMC requirements for electrical equipment for measurement, control, and laboratory use.
WaveSurfer 3000 Oscilloscopes EUROPEAN CONTACT: Teledyne LeCroy Europe GmbH Waldhofer Str 104 D-69123 Heidelberg Germany Tel: (49) 6221 82700 Australia & New Zealand Declaration of Conformity– EMC Oscilloscope complies with the EMC provision of the Radio Communications Act per the following standards, in accordance with requirements imposed by Australian Communication and Media Authority (ACMA): EN 55011:2010 Radiated and Conducted Emissions, Group 1, Class A, in accordance with EN61326-1:2013 and EN61326
Operator's Manual U.S. Nationally Recognized Agency Certification The oscilloscope has been certified by Underwriters Laboratories (UL) to conform to the following safety standard and bears UL Listing Mark: UL 61010-1 Third Edition – Safety standard for electrical measuring and test equipment. Canadian Certification The oscilloscope has been certified by Underwriters Laboratories (UL) to conform to the following safety standard and bears cUL Listing Mark: CAN/CSA-C22.2 No. 61010-1-12.
WaveSurfer 3000 Oscilloscopes Warranty THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE OR USE. TELEDYNE LECROY SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES, WHETHER IN CONTRACT OR OTHERWISE. THE CUSTOMER IS RESPONSIBLE FOR THE TRANSPORTATION AND INSURANCE CHARGES FOR THE RETURN OF PRODUCTS TO THE SERVICE FACILITY.