User manual
Operator's Manual
2. On the Digital<#> set up dialog, check the boxes for lines D0 through D15 that comprise the group.
Touch the Display D0-D7 and Display D8-D15 buttons to quickly turn on the entire digital bank, or touch
the Right and Left Arrow buttons to switch between each digital bank as you make line selections.
NOTE: Each group can consist of anywhere from 1 to 16 of the leads that are (or will be) connected to
signal, from either digital bank regardless of the Logic set on the bank. It does not matter if the some or
all of the lines have been included in other groups.
3. When all group members are selected, optionally rename them.
4. Go on to set up the digital display for the group. Check Group to enable the display.
5. When you're finished on the Digital<#> dialog, touch the Logic Setup tab and choose the Logic Family that
applies to each digital bank, or set custom Threshhold values.
Digital Display Set Up
You can choose the type and position of the digital traces that appear on screen for each digital group.
1. Set up the digital group.
2. Choose a Display Mode:
l Lines (default) shows a time-correlated trace indicating high, low, and transitioning points (relative
to the Threshold) for every digital line in the group. The size and placement of the lines depend on
the number of lines, the Vertical Position and Group Height settings.
l Bus collapses the lines in a group into their Hex values. It appears immediately below all the Line
traces when both are selected.
l Lines & Bus displays both line and bus traces at once.
3. In Vertical Position, enter the number of divisions (positive or negative) relative to the zero line of the grid
where the display begins.The top of the first trace appears at this position.
4. In Group Height, enter the total number of grid divisions the entire display should occupy. All the selected
traces (Line and Bus) will appear in this much space.
31