Press adjacent Operation buttons and use the Digital Input Front Panel
Controls (on page 27) to make specific adjustment/selection parameter
choices. For more information on getting to this point, see Navigating
Interfaces, Adjusting Parameters, and Making Selections (on page 16).
Mod → AM Operation menu selections and parameters include AM Freq,
AM Depth, Type, Shape, and Source as follows:
AM Freq - This parameter sets the modulating waveform
frequency. The internal source frequency range is 2 MHz to 20
kHz.
AM Depth - This parameter sets the amplitude range.
Note: Amplitude range is also referred to as amplitude depth or
percentage modulation. It's a percentage value that varies from
1 to 120% . When set to 0%, the output amplitude is
approximately half of the user set amplitude value.When set to
0%, the output amplitude is approximately half of the user set
amplitude value. When set to 100%, the output amplitude
matches the amplitude set. When using an external source, the
AM depth is controlled by the voltage level of the connector
attached to Modulation In on the Back Panel (on page 14). ±6V
corresponds to 100% of your current depth setting.
Type - Amplitude modulation.
Shape - Choose the waveform shape type (used as a carrier
waveform) for modulation. Sine, Square, Ramp, or Arb waveforms
may be modulated, while Pulse, Noise, and DC cannot.
Source - Select Internal or External. If External is selected, use the
Modulation In connector on the Back Panel (on page 14).
Generating a FM (Frequency Modulation)
Modulated Waveform
After selecting the Mod → FM modulated waveform operator button as
explained in the Generating Modulated Waveforms Overview (on page
41), the WaveStation shows the FM Operation menu and corresponding
Display and Detail dialogs.