70  WA1K2K-OM-E RevB 
PLEASE NOTE THE FOLLOWING:  
  The Recall option is excluded from Save/Recall menus when 
Picture or CSV is selected as the Type. 
  The Load option is disabled (grayed-out) on menus inside the 
SAVE ALL screen when Pictures or CSV file types are selected as 
Type. 
Recalling Factory Settings 
The oscilloscope is set up for normal operation when it is shipped from 
the factory. 
There are two ways to reload the factory settings: 
1.  Press the DEFAULT SETUP front panel button. 
OR 
2.  Press the Save/Recall front panel button. The SAVE/REC menu 
opens. Press the Type option button and select Factory. 
3.  Now, press the Load option button. 
Applying the default setup does not reset the following: 
  Language option 
  Saved reference waveform files 
  Saved setup files 
  Display contrast 
  Calibration data 